Shen L, Wang X N, Li D J, Wang Z Y, Chen W, Zhao Q, Li Y F, Cui R, Yang W L
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology & Visual Sciences Key Lab., Beijing 100730, China.
Zhonghua Yan Ke Za Zhi. 2018 Sep 11;54(9):678-682. doi: 10.3760/cma.j.issn.0412-4081.2018.09.008.
To explore the difference and agreement in the measurement of anterior chamber depth (ACD) and anterior chamber angle data in age-related cataract patients between swept source anterior segment optical coherence tomography (SS-AS-OCT) and ultrasound biomicroscopy (UBM). A prospective study. Forty-five eyes from 45 age-related cataract patients were selected at Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University from November to December 2016. There were 23 males and 22 females aged (51±19) years. SS-AS-OCT and UBM were used to measure the ACD, angle opening distance (AOD 500) and trabecular-iris angle (TIA 500). The paired samples -test and Pearson linear correlation were used for statistical analysis, and the Bland-Altman analysis was used to evaluate the measurement consistency of the two tools. The ACD measured by SS-AS-OCT and UBM was (2.750±0.544) mm and (2.775±0.531) mm, respectively, the difference was statistically significant (2.109, 0.040), and the measurement results were positively correlated (0.989, 0.001). In the nasal, temporal, upper and lower positions, the AOD 500 and TIA 500 values measured by SS-AS-OCT and UBM had a significant difference [AOD 500: (0.451±0.201) mm .(0.421±0.194) mm, (0.499±0.229) mm . (0.465±0.207) mm, (0.390±0.215) mm . (0.355±0.208) mm, (0.441±0.226) mm . (0.423±0.215) mm, 4.262, 4.350, 5.355, 4.415; TIA 500: 39.587°±14.698°. 36.944°±13.790°, 42.149°±14.138°. 40.289°±13.041°, 36.482°±15.887°. 35.600°±15.215°, 38.504°±14.805°. 36.818°±14.931°, 5.131, 4.658, 4.174, 5.731; all 0.001] and a significant linear correlation (AOD 500: 0.971, 0.975, 0.979, 0.973; TIA 500: 0.972, 0.957, 0.956, 0.970; all 0.001). The Bland-Altman method showed that the two modalities had comparable results for the ACD, AOD 500 and TIA 500 (2.22%-6.67% of the measured value outside the 95% consistency boundaries). SS-OCT and UBM show a good agreement in the measurement of ACD and anterior chamber angle data in age-related cataract patients, though there is difference between them in measurement results. .
探讨扫频源眼前节光学相干断层扫描(SS-AS-OCT)与超声生物显微镜(UBM)测量年龄相关性白内障患者前房深度(ACD)和前房角数据的差异及一致性。一项前瞻性研究。2016年11月至12月,在北京同仁医院北京同仁眼科中心选取45例年龄相关性白内障患者的45只眼。其中男性23例,女性22例,年龄(51±19)岁。采用SS-AS-OCT和UBM测量ACD、房角开放距离(AOD 500)和小梁-虹膜夹角(TIA 500)。采用配对样本t检验和Pearson线性相关进行统计分析,采用Bland-Altman分析评估两种工具的测量一致性。SS-AS-OCT和UBM测量的ACD分别为(2.750±0.544)mm和(2.775±0.531)mm,差异有统计学意义(t=2.109,P=0.040),测量结果呈正相关(r=0.989,P=0.001)。在鼻侧、颞侧(原文temporal误写为temporal,应改为temporal)、上方和下方位置,SS-AS-OCT和UBM测量的AOD 500和TIA 500值有显著差异[AOD 500:(0.451±0.201)mm.(0.421±0.194)mm,(0.499±0.229)mm.(0.465±0.207)mm,(0.390±0.215)mm.(0.355±0.208)mm,(0.441±