To test whether some biometry measurements provided by the Lenstar LS900 compared well with the AL-Scan, Pentacam rotating Scheimpflug camera, Ultrasound Biomicroscopy (UBM) and Tomey EM-3000.
Two hundred and one patients having routine cataract surgery had standard preoperative assessment. In this clinical study, the axis length (AL) and lens thickness (LT) were taken by Lenstar LS900 and AL-Scan; anterior chamber depth (ACD) was taken by Lenstar LS900, A-Scan, Pentacam and UBM; central corneal thickness (CCT) was taken by Lenstar LS900, Pentacam and Tomey EM-3000. The results were compared using a Wilcoxon-Mann-Whitney U test and Pearson correlation calculations. Agreement was assessed through intraclass correlation coefficients and Bland-Altman plots.
The highest correlation was found between Lenstar and AL-Scan for AL (r = 0.975; P < 0.001). For LT measurements, the correlation between these two devices was also good (r = 0.699; P < 0.001). Excellent correlations were showed between Lenstar and Pentacam or UBM for ACD (r = 0.948, 0.704, respectively, both P < 0.001), but not between Lenstar and AL-Scan (r = 0.453, P < 0.001). The correlations of CCT between Lenstar and Pentacam or Tomey EM-3000 were both excellent (r = 0.817, 0.882, respectively, both P < 0.001).
In phakic eyes of cataract patients, measurements of AL, LT, ACD and CCT from Lenstar LS900 yielded results that correlated very well with other clinical instruments.
测试 Lenstar LS900 提供的一些生物测量值与 AL-Scan、Pentacam 旋转式 Scheimpflug 相机、超声生物显微镜 (UBM) 和 Tomey EM-3000 的比较情况。
对 201 例行常规白内障手术的患者进行了标准术前评估。在这项临床研究中,Lenstar LS900 用于测量眼轴长度 (AL) 和晶状体厚度 (LT),AL-Scan 用于测量前房深度 (ACD);Lenstar LS900、A 扫描、Pentacam 和 UBM 用于测量中央角膜厚度 (CCT)。使用 Wilcoxon-Mann-Whitney U 检验和 Pearson 相关计算比较结果。通过组内相关系数和 Bland-Altman 图评估一致性。
Lenstar 和 AL-Scan 之间的 AL 相关性最高 (r=0.975;P<0.001)。对于 LT 测量值,这两种设备之间的相关性也很好 (r=0.699;P<0.001)。Lenstar 和 Pentacam 或 UBM 之间的 ACD 相关性极好 (r=0.948、0.704,均 P<0.001),但 Lenstar 和 AL-Scan 之间的相关性较差 (r=0.453,P<0.001)。Lenstar 和 Pentacam 或 Tomey EM-3000 之间的 CCT 相关性均极好 (r=0.817、0.882,均 P<0.001)。
在白内障患者的未植入眼,Lenstar LS900 测量的 AL、LT、ACD 和 CCT 值与其他临床仪器的测量结果相关性非常好。