Dada Tanuj, Sihota Ramanjit, Gadia Ritu, Aggarwal Anand, Mandal Subrata, Gupta Viney
Glaucoma Research Laboratory, Dr. Rajendra Prasad Centre for Ophthalmic Sciences, All India Institute of Medical Sciences, New Delhi, India.
J Cataract Refract Surg. 2007 May;33(5):837-40. doi: 10.1016/j.jcrs.2007.01.021.
To compare anterior segment parameters using quantitative imaging by anterior segment optical coherence tomography (AS-OCT) and ultrasound biomicroscopy (UBM).
Tertiary-care glaucoma research center.
Sixty-three eyes of 63 subjects had anterior segment evaluation by AS-OCT (Visante-Zeiss) and UBM (Paradigm). Central corneal thickness (CCT), anterior chamber depth (ACD) (measured from the central corneal endothelium to the anterior lens capsule), and the peripheral iridocorneal angles (temporal and nasal) were assessed and compared.
There was an excellent correlation between AS-OCT and UBM measurements for the nasal angle (r = 0.84; P<.0001), temporal angle (r = 0.86; P<.0001), ACD (r = 0.97; P<.0001), and CCT (r = 0.91; P<.0001). There was no significant difference (paired t test) between the mean ACD, CCT, and angle parameters measured by AS-OCT or UBM. The mean values of the parameters measured by AS-OCT and UBM were, respectively, as follows: nasal angle, 26.25 degrees +/- 11.0 (SD) and 28.27 +/- 11.3 degrees (P = .3); temporal angle, 25.1 +/- 11.4 degrees and 28.3 +/- 13.5 degrees (P = .15); ACD, 2.85 +/- 0.5 mm and 2.78 +/- 0.5 mm (P = .2); and CCT, 512 +/- 46 microm and 502 +/- 46 microm (P = .25). The AS-OCT images showed sharper definition of the scleral spur than the UBM images.
Anterior segment optical coherence tomography and UBM can both be used for anterior segment measurements and yielded comparable results.
通过眼前节光学相干断层扫描(AS-OCT)和超声生物显微镜检查(UBM)的定量成像来比较眼前节参数。
三级医疗青光眼研究中心。
63名受试者的63只眼睛接受了AS-OCT(Visante-蔡司)和UBM(Paradigm)的眼前节评估。评估并比较中央角膜厚度(CCT)、前房深度(ACD)(从中央角膜内皮测量至晶状体前囊)以及周边虹膜角膜角(颞侧和鼻侧)。
鼻侧角(r = 0.84;P <.0001)、颞侧角(r = 0.8...