Velez-Quinones Maria A, Xu Hanna, Vo Nhu, Gaitens Joanna M, McDiarmid Melissa A, Lewin-Smith Michael R, Strathmann Frederick G
Environmental Toxicology Laboratory, The Joint Pathology Center, Silver Spring, Maryland.
Division of Occupational and Environmental Medicine, Department of Medicine, University of Maryland School of Medicine, Baltimore, Maryland.
Curr Protoc Toxicol. 2018 Nov;78(1):e59. doi: 10.1002/cptx.59. Epub 2018 Oct 4.
We have developed and validated a method for the simultaneous quantitative measurement of total uranium (TU) and uranium U/ U isotopic ratio (UIR) in urine by inductively coupled plasma mass spectrometry (ICP-MS) using a Thermo Scientific iCAP-Q instrument. The performance characteristics of the assay were determined to be in compliance with clinical laboratory standards. The assay was linear in the concentration range of 1.0 to 500.0 ng/liter TU. The method was precise and accurate with limits of detection of 2.5 ng/liter for TU and 9.8 ng/liter for UIR. The accuracy was >93% and the coefficient of variation (% CV) was <5.0% for both TU and UIR. All results were within established guidelines and agreed-upon criteria, and the results fell within the certified range for the reference controls. The method has thus been shown to be effective as a simple, precise, and sensitive analytical technique for testing urine samples. © 2018 by John Wiley & Sons, Inc.
我们开发并验证了一种使用赛默飞世尔科技的iCAP-Q仪器,通过电感耦合等离子体质谱法(ICP-MS)同时定量测量尿液中总铀(TU)和铀U/U同位素比率(UIR)的方法。测定的性能特征经确定符合临床实验室标准。该测定在1.0至500.0纳克/升TU的浓度范围内呈线性。该方法精确且准确,TU的检测限为2.5纳克/升,UIR的检测限为9.8纳克/升。TU和UIR的准确度均>93%,变异系数(%CV)均<5.0%。所有结果均在既定指南和商定标准范围内,且结果落在参考对照的认证范围内。因此,该方法已被证明是一种用于检测尿液样本的简单、精确且灵敏的有效分析技术。© 2018约翰威立父子出版公司