Xiao Ge, Jones Robert L, Saunders David, Caldwell Kathleen L
Inorganic and Radiation Analytical Toxicology Branch, Centers for Disease Control and Prevention, 4770 Buford HWY, Mail Stop F50, Atlanta, GA 30341, USA
Inorganic and Radiation Analytical Toxicology Branch, Centers for Disease Control and Prevention, 4770 Buford HWY, Mail Stop F50, Atlanta, GA 30341, USA.
Radiat Prot Dosimetry. 2014 Dec;162(4):618-24. doi: 10.1093/rpd/ncu023. Epub 2014 Feb 20.
Quantification of the isotopic composition of uranium in urine at low levels of concentration is important for assessing both military and civilian populations' exposures to uranium. However, until now there has been no convenient, precise method established for rapid determination of multiple uranium isotope ratios. Here, the authors report a new method to measure (234)U/(238)U, (235)U/(238)U and (236)U/(238)U. It uses solid-phase chelation extraction (via TRU columns) of actinides from the urine matrix, followed by measurement using a magnetic sector field inductively coupled plasma mass spectrometer (SF-ICP-MS-Thermo Element XR) equipped with a high-efficiency nebulizer (Apex PFA microflow) and coupled with a membrane desolvating nebulizer system (Aridus II™). This method provides rapid and reliable results and has been used successfully to analyse Certified Reference Materials.
对低浓度尿液中铀的同位素组成进行定量分析,对于评估军事人员和平民的铀暴露情况都很重要。然而,迄今为止,尚未建立一种方便、精确的方法来快速测定多种铀同位素比率。在此,作者报告了一种测量(234)U/(238)U、(235)U/(238)U和(236)U/(238)U的新方法。该方法采用从尿液基质中通过TRU柱进行锕系元素的固相螯合萃取,随后使用配备高效雾化器(Apex PFA微流雾化器)并与膜去溶剂化雾化器系统(Aridus II™)联用的磁扇形场电感耦合等离子体质谱仪(SF-ICP-MS-Thermo Element XR)进行测量。该方法能提供快速且可靠的结果,并已成功用于分析有证标准物质。