Automated gonioscopy photography for iridocorneal angle grading.

作者信息

Teixeira Filipa, Sousa David C, Leal Inês, Barata André, Neves Carlos M, Pinto Luís A

机构信息

Department of Ophthalmology, Hospital de Santa Maria, Lisbon, Portugal.

Clínica Universitária de Oftalmologia, Faculdade de Medicina, Universidade de Lisboa, Lisbon, Portugal.

出版信息

Eur J Ophthalmol. 2020 Jan;30(1):112-118. doi: 10.1177/1120672118806436. Epub 2018 Oct 26.

Abstract

PURPOSE

The aim of this study is to assess the agreement between manual and automated gonioscopy for iridocorneal angle opening.

METHODS

The research is a cross-sectional observational study. Manual and automated gonioscopy were performed to consecutive patients in a glaucoma clinic. Iridocorneal angle opening grading was performed according to Shaffer's classification. Automated gonioscopy was performed with NGS-1 automated gonioscope (NIDEK Co., Gamagori, Japan). The automated gonio-photos were graded by two independent observers. Agreement between automated and manual gonioscopy and also among raters was ascertained by Fleiss' kappa statistic and comparison of area under curve.

RESULTS

In total, 88 eyes of 47 subjects were analysed. Mean age was 63 ± 10 years. Twenty eyes (22.7%) were excluded from grading due to poor quality images. Angle closure was detected in 23.4% with dynamic gonioscopy in comparison with 4.3% using automated image grading. The agreement for angle closure diagnosis between dynamic and automated gonioscopy was low (κ = 0.09 ± 0.10; p = 0.18). The area under curve for detecting eyes with angle closure showed poor accuracy between automated and manual methods (area under curve: 0.53 ± 0.05, 95% confidence interval: 0.44-0.62). There was modest inter-rater agreement for angle opening assessment of automated images with Fleiss' kappa of 0.17 (95% confidence interval: 0.035-0.238).

CONCLUSION

Manual and automated gonioscopy showed only slight agreement for the assessment of iridocorneal angle opening status. Further improvements of the NGS-1 automated gonioscopy and technique are desired for widespread use in a real-life setting.

摘要

目的

本研究旨在评估手动房角镜检查与自动房角镜检查在虹膜角膜角开放度方面的一致性。

方法

本研究为横断面观察性研究。在青光眼诊所对连续的患者进行手动和自动房角镜检查。根据Shaffer分类法对虹膜角膜角开放度进行分级。使用NGS - 1自动房角镜(日本NIDEK公司,蒲郡市)进行自动房角镜检查。由两名独立观察者对自动房角镜照片进行分级。通过Fleiss卡方统计量和曲线下面积比较来确定自动房角镜检查与手动房角镜检查之间以及评估者之间的一致性。

结果

共分析了47名受试者的88只眼。平均年龄为63±10岁。由于图像质量差,20只眼(22.7%)被排除在分级之外。动态房角镜检查发现23.4%的患者存在房角关闭,而自动图像分级法检测到的比例为4.3%。动态房角镜检查与自动房角镜检查在房角关闭诊断方面的一致性较低(κ = 0.09±0.10;p = 0.18)。自动和手动方法在检测房角关闭眼方面的曲线下面积显示准确性较差(曲线下面积:0.53±0.05,95%置信区间:0.44 - 0.62)。对于自动图像的房角开放度评估,评估者之间的一致性一般,Fleiss卡方值为0.17(95%置信区间:0.035 - 0.238)。

结论

手动房角镜检查与自动房角镜检查在评估虹膜角膜角开放状态方面仅显示出轻微的一致性。为了在实际应用中广泛使用,需要对NGS - 1自动房角镜检查及其技术进行进一步改进。