根据 NEMA NU2-2012 标准对 Vereos PET/CT 系统进行性能评估。

Performance Evaluation of the Vereos PET/CT System According to the NEMA NU2-2012 Standard.

机构信息

QIMP Team, Center of Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria

Department of Medical Physics and Radiation Protection, Hospital Santa Creu i Sant Pau, Barcelona, Catalonia/Spain; and.

出版信息

J Nucl Med. 2019 Apr;60(4):561-567. doi: 10.2967/jnumed.118.215541. Epub 2018 Oct 25.

Abstract

The aim of this study was to evaluate the physical performance of the Vereos whole-body PET/CT system according to the National Electrical Manufacturers Association (NEMA) NU2-2012 standard and to compare it with other state-of-the-art PET/CT systems. Spatial resolution, sensitivity, count-rate performance, count rate accuracy, and image quality were assessed. Specifically, spatial resolution was measured using an F point-source. System sensitivity was calculated from acquisitions of an F line source inside aluminum tubes of varying thickness. Assessment of count rate performance and count rate accuracy was based on measurements of an F line source inside a 20-cm-diameter polyethylene cylinder. PET image quality was assessed using a NEMA IQ phantom. All measurements were performed according to the predefined and implemented NEMA NU2-2012 acquisition protocols at a clinical installation of a Vereos PET/CT system. Evaluation was performed using the software provided by the vendor. The average (radial and tangential) transverse spatial resolution was 4.2, 4.5, and 5.5 mm in full width at half maximum for a 1-, 10-, and 20-cm radial offset, respectively, from the center of the field of view. Axial spatial resolution varied between 4.2 and 4.6 mm in full width at half maximum, depending on the radial source position. The average sensitivity was 5.2 cps/kBq. A peak noise-equivalent count (NEC) rate of 153.4 kcps was measured at an activity concentration of 54.9 kBq/mL. The scatter fraction at peak NEC rate was 33.9%, and the maximum count rate error at and below peak NEC rate was 6.8%. Contrast recovery coefficients varied from 54.3% (10-mm sphere) to 83.9% (22-mm sphere) for the hot spheres, and between 81.4% (28-mm sphere) and 87% (37-mm sphere) for the cold spheres for a given sphere-to-background ratio of 4:1. The overall performance characteristics of the Vereos PET/CT system are comparable to state-of-the-art whole-body PET/CT systems with the exception that the peak NEC rate occurs at a higher activity concentration, thus indicating the ability of the Vereos PET/CT system to cover a wider range of activities.

摘要

这项研究的目的是根据美国电器制造商协会(NEMA)NU2-2012 标准评估 Vereos 全身 PET/CT 系统的物理性能,并将其与其他最先进的 PET/CT 系统进行比较。评估了空间分辨率、灵敏度、计数率性能、计数率准确性和图像质量。具体来说,空间分辨率是使用 F 点源测量的。系统灵敏度是通过在不同厚度的铝管内采集 F 线源计算得出的。计数率性能和计数率准确性的评估是基于在 20 厘米直径的聚乙烯圆柱内测量 F 线源得出的。使用 NEMA IQ 体模评估 PET 图像质量。所有测量均根据 Vereos PET/CT 系统临床安装的预定义和实施的 NEMA NU2-2012 采集协议进行。评估使用供应商提供的软件进行。在视野中心的 1、10 和 20 厘米径向偏移处,1、10 和 20 厘米处的平均(径向和切向)横向空间分辨率分别为 4.2、4.5 和 5.5 毫米。轴向空间分辨率在半最大值全宽处介于 4.2 和 4.6 毫米之间,具体取决于径向源位置。平均灵敏度为 5.2 cps/kBq。在 54.9 kBq/mL 的活度浓度下,测量到峰值噪声等效计数(NEC)率为 153.4 kcps。在峰值 NEC 率时的散射分数为 33.9%,在峰值 NEC 率以下的最大计数率误差为 6.8%。对于给定的 4:1 球与背景比,热球的对比恢复系数从 54.3%(10 毫米球)变化到 83.9%(22 毫米球),冷球的对比恢复系数在 81.4%(28 毫米球)和 87%(37 毫米球)之间变化。Vereos PET/CT 系统的整体性能特征与最先进的全身 PET/CT 系统相当,不同之处在于峰值 NEC 率出现在更高的活度浓度下,这表明 Vereos PET/CT 系统能够覆盖更宽的活度范围。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验