Santos García Alba, Macías Carmona Nicolás, Vega Martínez Almudena, Abad Estébanez Soraya, Linares Grávalos Tania, Aragoncillo Sauco Inés, Verdalles Guzmán Ursula, Panizo González Nayara, Cruzado Vega Leonidas, López Gómez Juan Manuel
Nephrology Department, Hospital Gregorio Marañón, Madrid, Spain.
Nephrology Department, Hospital Universitario de Elche, Alicante, Spain.
Hemodial Int. 2019 Jan;23(1):50-57. doi: 10.1111/hdi.12685. Epub 2018 Oct 27.
The aim of this study is to compare molecule removal and albumin leakage in postdilution online hemodiafiltration with different high-flux dialyzers.
We studied seven high-flux dialyzers (Polyflux 210H®, Evodial 2.2®, FxCordiax1000®, Elisio21H®, TS-2.1SL®, XevontaHi20®, VitaPES 210-HF®) in 6 patients. The reduction ratio (RR) of small- and middle-sized molecules was calculated. Dialysate samples were collected to estimate the albumin leakage.
Global differences between dialyzers were observed in the RR of ß2 microglobulin (P =0.003) and prolactin (P =0.013). The mean loss of albumin in the dialysate per session varied between 114 ± 67 mg (with Evodial 2.2) and 2621 ± 1363 mg per session (with XevontaHi20). We found global differences between dialyzers in total albumin loss (P = 0.05).
We demonstrated that the performance of high-flux dialyzers was different among the types and that not all high-flux dialyzers should be considered equal.
本研究旨在比较不同高通量透析器在后置稀释在线血液透析滤过中的分子清除率及白蛋白漏出情况。
我们对6例患者使用7种高通量透析器(Polyflux 210H®、Evodial 2.2®、FxCordiax1000®、Elisio21H®、TS - 2.1SL®、XevontaHi20®、VitaPES 210 - HF®)进行研究。计算小分子和中分子的清除率(RR)。收集透析液样本以评估白蛋白漏出情况。
观察到不同透析器之间在β2微球蛋白清除率(P = 0.003)和催乳素清除率(P = 0.013)方面存在总体差异。每次透析过程中透析液中白蛋白的平均丢失量在114±67毫克(使用Evodial 2.2时)至每次2621±1363毫克(使用XevontaHi20时)之间。我们发现不同透析器在白蛋白总丢失量方面存在总体差异(P = 0.05)。
我们证明了不同类型的高通量透析器性能不同,并非所有高通量透析器都应被视为等同。