Ma Jack X, Wang Li, Weikert Mitchell P, Montes de Oca Ildamaris, Koch Douglas D
McGovern Medical School, Houston, TX.
Department of Ophthalmology, Cullen Eye Institute, Baylor College of Medicine, Houston, TX.
Cornea. 2019 Jan;38(1):67-73. doi: 10.1097/ICO.0000000000001806.
To evaluate the repeatability and between-machine/operator reproducibility of optical coherence tomography (RTVue-XR) for mapping corneal epithelial thickness (ET) and corneal thickness (CT) over a 9-mm zone.
In 5 groups of eyes (normal, contact lens, dry eye, post-LASIK/PRK, and keratoconus), corneal ET and CT were evaluated in 25 areas in the following 4 regions: 1) 1 central zone within 0 to 2.0 mm diameter, 2) 8 paracentral zones from 2.0 to 5.0 mm diameter, 3) 8 midperipheral zones from 5.0 to 7.0 mm diameter, and 4) 8 peripheral zones from 7.0 to 9.0 mm diameter. Three devices with designated operators were used to obtain 3 scans per device for each eye. The repeatability, reproducibility, and coefficient of variation were calculated.
For all areas up to 9 mm diameter, in normal eyes, contact lens eyes, dry eyes, post- LASIK/PRK eyes, and keratoconic eyes, repeatability and reproducibility ranged from 0.7 to 1.5, 1.0 to 2.6, 0.9 to 2.0, 0.7 to 2.5, and 1.1 to 2.3 μm for ET and 1.5 to 8.3, 2.3 to 12.1, 1.7 to 14.2, 1.4 to 13.7, and 6.9 to 18.9 μm for CT, respectively. The ranges of coefficient of variation values were 1.2% to 3.0%, 1.9% to 5.5%, 1.8% to 4.3%, 1.3% to 5.2%, and 2.2% to 4.5% for ET and 0.3% to 1.2%, 0.4% to 1.8%, 0.4% to 2.1%, 0.3% to 2.0%, and 1.2% to 3.0% for CT, respectively.
Optical coherence tomography produced excellent repeatability and reproducibility for corneal ET and CT measurements up to a 9-mm zone in normal eyes and eyes with different corneal conditions. The CT measurements in keratoconic eyes were more variable than those in eyes with other corneal conditions.
评估光学相干断层扫描(RTVue-XR)在绘制9毫米区域角膜上皮厚度(ET)和角膜厚度(CT)时的可重复性以及不同机器/操作人员之间的再现性。
在5组眼睛(正常眼、戴隐形眼镜眼、干眼、准分子激光原位角膜磨镶术/准分子激光角膜切削术后眼和圆锥角膜眼)中,在以下4个区域的25个部位评估角膜ET和CT:1)直径0至2.0毫米范围内的1个中央区;2)直径2.0至5.0毫米的8个旁中央区;3)直径5.0至7.0毫米的8个中周边区;4)直径7.0至9.0毫米的8个周边区。使用三台设备并指定操作人员,对每只眼睛的每台设备进行3次扫描。计算可重复性、再现性和变异系数。
对于直径达9毫米的所有区域,在正常眼、戴隐形眼镜眼、干眼、准分子激光原位角膜磨镶术/准分子激光角膜切削术后眼和圆锥角膜眼中,ET的可重复性和再现性范围分别为0.7至1.5、1.0至2.6、0.9至2.0、0.7至2.5和1.1至2.3微米,CT的可重复性和再现性范围分别为1.5至8.3、2.3至12.1、1.7至14.2、1.4至13.7和6.9至18.9微米。ET的变异系数值范围分别为1.2%至3.0%、1.9%至5.5%、1.8%至4.3%、1.3%至5.2%和2.2%至4.5%,CT的变异系数值范围分别为0.3%至1.2%、0.4%至1.8%、0.4%至2.1%、0.3%至2.0%和1.2%至3.0%。
光学相干断层扫描在正常眼和不同角膜状况的眼中,对9毫米区域内的角膜ET和CT测量具有出色的可重复性和再现性。圆锥角膜眼中的CT测量比其他角膜状况的眼中更具变异性。