Department of Radiology, Ajou University School of Medicine, Ajou University Hospital, Suwon, South Korea.
Department of Gastroenterology, Ajou University School of Medicine, Ajou University Hospital, Suwon, South Korea.
J Magn Reson Imaging. 2019 Jul;50(1):305-314. doi: 10.1002/jmri.26575. Epub 2018 Nov 14.
Complex-based chemical shift imaging-based magnetic resonance imaging (CSE-MRI) is emerging as a preferred method for noninvasively quantifying proton density fat fraction (PDFF), a promising quantitative imaging biomarker (QIB) for longitudinal hepatic steatosis measurement.
To determine linearity, bias, repeatability, and reproducibility of the PDFF measurement using CSE-MRI (CSE-PDFF) across scan intervals, MR field strengths, and readers in phantom and nonalcoholic fatty liver disease (NAFLD) patients.
Institutional Review Board (IRB)-approved prospective.
Fat-water phantom and 20 adult patients.
FIELD STRENGTH/SEQUENCE: 1.5 T and 3.0 T MR systems and a commercially available CSE-MRI sequence (IDEAL-IQ).
Two independent readers measured CSE-PDFF of fat-water phantom and NAFLD patients across two field strengths and scan intervals (same-day and 2-week) each and in a combination of both. MR spectroscopy-based PDFF (MRS-PDFF) was used as the reference standard for phantom PDFF.
Linearity and bias of measurement were evaluated by linear regression analysis and Bland-Altman plots, respectively. Repeatability and reproducibility were assessed by coefficient of variance and repeatability / reproducibility coefficients (RC). The intraclass correlation coefficient was used to validate intra- and interobserver agreements.
CSE-PDFF showed high linearity and small bias (-0.6-0.4 PDFF%) with 95% limits of agreement within ±2.9 PDFF% across field strengths, 2-week interscan period, and readers in the clinical scans. CSE-PDFF was highly repeatable and reproducible both in phantom and clinical scans, with the largest observed RC across field strengths and 2-week interscan period being 3 PDFF%.
CSE-PDFF is a robust QIB with high linearity, small bias, and excellent repeatability/reproducibility. A change of more than 3 PDFF% across field strengths within 2 weeks of scan interval likely reflects a true change, which is well within the clinically acceptable range.
3 Technical Efficacy: Stage 2 J. Magn. Reson. Imaging 2019;50:305-314.
基于复杂的化学位移成像的磁共振成像(CSE-MRI)正在成为一种非侵入性定量质子密度脂肪分数(PDFF)的首选方法,PDFF 是一种有前途的用于肝脂肪变性纵向测量的定量成像生物标志物(QIB)。
在体模和非酒精性脂肪性肝病(NAFLD)患者中,确定 CSE-MRI(CSE-PDFF)在扫描间隔、MR 场强和读者之间的 PDFF 测量的线性、偏差、重复性和再现性。
机构审查委员会(IRB)批准的前瞻性研究。
脂肪-水体模和 20 名成年患者。
场强/序列:1.5T 和 3.0T MR 系统和一种商业上可用的 CSE-MRI 序列(IDEAL-IQ)。
两位独立的读者在两个场强和扫描间隔(当天和 2 周)以及两者的组合中测量了脂肪-水体模和 NAFLD 患者的 CSE-PDFF,并使用基于磁共振波谱的 PDFF(MRS-PDFF)作为体模 PDFF 的参考标准。
通过线性回归分析和 Bland-Altman 图分别评估测量的线性和偏差。通过方差系数和重复性/再现性系数(RC)评估重复性和再现性。使用组内相关系数验证观察者内和观察者间的一致性。
在临床扫描中,CSE-PDFF 在整个场强、2 周的扫描间隔和读者中均表现出高度的线性和较小的偏差(-0.6-0.4 PDFF%),95%一致性区间在±2.9 PDFF%范围内。CSE-PDFF 在体模和临床扫描中均具有高度的重复性和再现性,在整个场强和 2 周的扫描间隔中,最大的观察到 RC 为 3 PDFF%。
CSE-PDFF 是一种具有高线性、小偏差和优异重复性/再现性的强大 QIB。在 2 周的扫描间隔内,场强变化超过 3 PDFF%可能反映了真实的变化,这在临床可接受的范围内。
3 技术功效:第 2 阶段 J. Magn. Reson. Imaging 2019;50:305-314.