Meier D, Maier S, Bösiger P
Institute of Biomedical Engineering, University of Zurich, Switzerland.
Magn Reson Med. 1988 Sep;8(1):25-34. doi: 10.1002/mrm.1910080104.
A broad spectrum of MR methods has been published in the last few years for the visualization and the quantification of flow in human blood vessels. We describe the results of an extended study to measure the flow quantitatively using the MR phase modulation method. A gradient echo sequence which allows short repetition times was synchronized with the ECG. The instantaneous velocity profiles and the integrated flow rate can be determined throughout the cardiac cycle in up to 45 consecutive time intervals. The instantaneous velocity distributions can be displayed as 2D profiles over the lumen of the blood vessel. The technique has been successfully evaluated with a set of phantom experiments. Following this, quantifications of the blood flow rates in vivo were performed on the abdominal aortas of healthy volunteers. The MR flow results were compared to those obtained with conventional Doppler ultrasound examinations. The results of both methods are in good quantitative agreement.
在过去几年里,已经发表了各种各样的磁共振(MR)方法,用于可视化和量化人体血管中的血流。我们描述了一项扩展研究的结果,该研究使用MR相位调制方法对血流进行定量测量。一个允许短重复时间的梯度回波序列与心电图(ECG)同步。在整个心动周期中,最多可在45个连续的时间间隔内确定瞬时速度剖面和积分流量。瞬时速度分布可以显示为血管腔内的二维剖面。该技术已通过一组体模实验成功评估。在此之后,对健康志愿者的腹主动脉进行了体内血流率的定量分析。将MR血流结果与传统多普勒超声检查获得的结果进行了比较。两种方法的结果在定量上吻合良好。