Łabuz Grzegorz, Papadatou Eleni, Khoramnia Ramin, Auffarth Gerd U
J Refract Surg. 2018 Dec 1;34(12):832-838. doi: 10.3928/1081597X-20181108-01.
To measure longitudinal chromatic aberration (LCA) of various intraocular lenses (IOLs) and to assess LCA effects on polychromatic image quality with a focus on multifocal IOL designs.
The LCA values of four multifocal IOL models (three diffractive models: AT LARA 829MP and AT LISA 809M [both from Carl Zeiss Meditec AG, Jena, Germany] and Restor SN6AD1 [Alcon Laboratories, Inc., Fort Worth, TX] and one refractive model [Mini Well Ready; SIFI Medtech, Sant'Antonio, Italy]) were compared with that of their monofocal counterparts. Optical properties were assessed using an optical bench device with spectral filters. LCA was calculated as a difference of a lens power at 480 and 644 nm. The optical quality was evaluated objectively by means of modulation transfer function metrics.
At far, LCA of an aphakic model eye (1.04 diopters [D]) increased with the AT Lisa, Restor, and Mini Well IOLs to 1.40, 1.91, and 1.27 D, respectively, but decreased to 0.78 D with the AT Lara IOL. The AT Lisa and Restor IOLs showed comparable results with their monofocal platforms. A near-focus LCA decreased only in the diffractive IOLs. At far, the polychromatic modulation transfer function was reduced in all IOLs, but LCA effects were attenuated at near.
Multifocal IOLs and their monofocal counterparts show similar material dispersion. Multifocal diffractive IOLs proved effective in reducing LCA, but the efficiency of the LCA correction differed depending on the optical design. LCA adversely affects the polychromatic image quality. [J Refract Surg. 2018;34(12):832-838.].
测量各种人工晶状体(IOL)的纵向色差(LCA),并评估LCA对多色图像质量的影响,重点关注多焦点IOL设计。
将四种多焦点IOL模型(三种衍射模型:AT LARA 829MP和AT LISA 809M[均来自德国耶拿的卡尔蔡司医疗技术公司]以及Restor SN6AD1[爱尔康实验室公司,得克萨斯州沃思堡]和一种折射模型[Mini Well Ready;意大利圣安东尼奥的SIFI Medtech])的LCA值与其单焦点对应物进行比较。使用带有光谱滤光片的光具座设备评估光学性能。LCA计算为480和644nm处镜片屈光力的差值。通过调制传递函数指标客观评估光学质量。
在远视力时,无晶状体模型眼的LCA(1.04屈光度[D])随着AT Lisa、Restor和Mini Well IOL分别增加到1.40、1.91和1.27 D,但使用AT Lara IOL时降至0.78 D。AT Lisa和Restor IOL与其单焦点平台显示出可比的结果。近焦点LCA仅在衍射IOL中降低。在远视力时,所有IOL的多色调制传递函数均降低,但LCA效应在近视力时减弱。
多焦点IOL及其单焦点对应物显示出相似的材料色散。多焦点衍射IOL被证明在减少LCA方面有效,但LCA校正的效率因光学设计而异。LCA对多色图像质量有不利影响。[《屈光手术杂志》。2018;34(12):832 - 838。]