Suppr超能文献

心脏正电子发射断层扫描和单光子发射计算机断层心肌灌注成像的患者辐射暴露现状。

Current Status of Patient Radiation Exposure of Cardiac Positron Emission Tomography and Single-Photon Emission Computed Tomographic Myocardial Perfusion Imaging.

机构信息

Department of Cardiovascular Medicine, Morristown Medical Center, NJ (M.C.D., G.V.H.).

Waterbury Hospital, CT (J.B.L.).

出版信息

Circ Cardiovasc Imaging. 2018 Dec;11(12):e007565. doi: 10.1161/CIRCIMAGING.118.007565.

Abstract

BACKGROUND

Radiation exposure during nuclear cardiology procedures has received much attention and has prompted citations for radiation reduction. In 2010, the American Society of Nuclear Cardiology recommended reducing the average patient study radiation exposure to <9 mSv in 50% of studies by 2014. Cardiac positron emission tomography (PET) for myocardial perfusion imaging (MPI) has emerged within recent years, but current radiation exposure in cardiac nuclear PET laboratories is unknown. This study evaluated current reported patient radiation exposure from nuclear laboratories in the United States applying for Intersocietal Accreditation Commission accreditation for MPI using single photon emission computed tomography (SPECT) or PET.

METHODS AND RESULTS

This was an analysis of nuclear cardiology studies submitted to the Intersocietal Accreditation Commission for either or both cardiac PET and SPECT accreditation. Cardiac SPECT data represented year 2015 while PET data combined years 2013 to 2015. Data was analyzed with χ and Mann-Whitney U tests (reported as median, 25th percentile, and 75th percentile). Reported PET MPI radiation exposure for 111 laboratories (532 patient cases) was 3.7 (3.2-4.1) mSv per study with no geographic variation. Reported SPECT MPI radiation exposure for 665 laboratories (3067 patient studies) was 12.8 (12.2-14.3) mSv. Highest radiation exposure was found in the South region. Technetium-only studies resulted in a median of 12.2 mSv per study.

CONCLUSIONS

Radiation exposure from cardiac PET MPI in US laboratories applying for Intersocietal Accreditation Commission accreditation is low (111 laboratories, 3.7 mSv) and substantially lower than cardiac SPECT (665 laboratories, 12.8 mSv).

摘要

背景

核心脏病学检查过程中的辐射暴露引起了广泛关注,并促使人们采取措施减少辐射。2010 年,美国核心血管学会建议到 2014 年,将 50%的患者检查辐射暴露量减少到<9mSv。近年来,心脏正电子发射断层扫描(PET)心肌灌注成像(MPI)开始出现,但目前尚不清楚心脏核医学 PET 实验室的当前辐射暴露情况。本研究评估了目前在美国申请国际联合委员会 MPI 认证的核医学实验室报告的患者辐射暴露情况,这些实验室使用单光子发射计算机断层扫描(SPECT)或 PET。

方法和结果

这是对向国际联合委员会申请心脏 PET 和 SPECT 认证的核心脏病学研究的分析。心脏 SPECT 数据代表 2015 年,而 PET 数据则结合了 2013 年至 2015 年的数据。采用卡方和曼-惠特尼 U 检验(以中位数、25 百分位数和 75 百分位数报告)进行数据分析。报告的 111 个实验室(532 例患者)的 PET MPI 辐射暴露为 3.7(3.2-4.1)mSv/次,无地域差异。报告的 665 个实验室(3067 例患者)的 SPECT MPI 辐射暴露为 12.8(12.2-14.3)mSv。辐射暴露最高的是南部地区。锝单光子研究的中位数为 12.2mSv/次。

结论

在美国申请国际联合委员会认证的核医学实验室进行心脏 PET MPI 的辐射暴露较低(111 个实验室,3.7mSv),明显低于心脏 SPECT(665 个实验室,12.8mSv)。