Measurement of Dynamic Hyperinflation During the 6-Minute Walk Test Using a Mobile Device.

机构信息

Department of Pulmonary Medicine, Zuyderland Medical Center, Heerlen, The Netherlands.

Department of Pulmonary Medicine, University Medical Center Utrecht, Heidelberglaan, Utrecht, The Netherlands.

出版信息

Respir Care. 2019 Feb;64(2):182-188. doi: 10.4187/respcare.06307. Epub 2018 Dec 24.

Abstract

BACKGROUND

Dynamic hyperinflation is a major cause of dyspnea in patients with COPD. Dynamic hyperinflation is usually measured with cardiopulmonary exercise testing, but this test requires maximal effort from the patient, and inspiratory capacity (IC) measurements may influence test results. This study investigated another approach by measuring dynamic hyperinflation during the more natural and better-tolerated 6-min walk test (6MWT), using a mobile system for IC measurements.

METHODS

Twenty subjects were studied, including 10 healthy subjects and 10 subjects with COPD. Each subject performed a pair of 6MWTs.

RESULTS

The study showed that test-retest reliability in healthy subjects was excellent (intraclass correlation coefficient 0.995). In addition, mean resting IC in healthy subjects (3.27 ± 0.77 L) was not affected at the end of the 6MWT, whereas mean resting IC in subjects with COPD (2.45 ± 0.90 L) decreased to 1.93 L ± 0.68 L at the end of the 6MWT ( = .004). The tolerability of the device for subjects with COPD appeared to be only slightly hampered by an increased sense of dyspnea attributed to wearing a mask.

CONCLUSIONS

This study provides a proof of principle for mobile IC measurements during the 6MWT, demonstrating good test performance, reasonable tolerability, and a power to differentiate normal from pathologic conditions.

摘要

背景

动态过度充气是 COPD 患者呼吸困难的主要原因。动态过度充气通常通过心肺运动测试来测量,但该测试需要患者付出最大努力,并且吸气量(IC)测量可能会影响测试结果。本研究通过使用移动系统进行 IC 测量,在更自然和更耐受的 6 分钟步行测试(6MWT)中测量动态过度充气,从而提供了另一种方法。

方法

研究了 20 名受试者,包括 10 名健康受试者和 10 名 COPD 患者。每位受试者都进行了一对 6MWT。

结果

研究表明,健康受试者的测试-重测可靠性非常好(组内相关系数 0.995)。此外,健康受试者的静息 IC 平均值(3.27 ± 0.77 L)在 6MWT 结束时没有受到影响,而 COPD 患者的静息 IC 平均值(2.45 ± 0.90 L)在 6MWT 结束时下降至 1.93 L ± 0.68 L(P <.004)。对 COPD 患者而言,该设备的耐受性似乎仅因佩戴口罩引起的呼吸困难感略有增加而受到轻微阻碍。

结论

这项研究为 6MWT 期间的移动 IC 测量提供了原理证明,证明了良好的测试性能、合理的耐受性以及区分正常和病理情况的能力。