两种扫频源光学相干断层血管成像设备的脉络膜新生血管区和血管密度比较。

CHOROIDAL NEOVASCULAR AREA AND VESSEL DENSITY COMPARISON BETWEEN TWO SWEPT-SOURCE OPTICAL COHERENCE TOMOGRAPHY ANGIOGRAPHY DEVICES.

机构信息

Department of Ophthalmology, University Paris Est Creteil, Centre Hospitalier Intercommunal de Creteil, Creteil, France.

Department of Ophthalmology, University Vita-Salute, IRCCS Ospedale San Raffaele, Milan, Italy; and.

出版信息

Retina. 2020 Mar;40(3):521-528. doi: 10.1097/IAE.0000000000002430.

Abstract

PURPOSE

To compare choroidal neovascularization (CNV) area and vessel density (VD) measurements between two different swept-source optical coherence tomography angiography (SS-OCTA) devices.

METHODS

En face optical coherence tomography angiography (OCTA) images of patients affected by neovascular age-related macular degeneration were collected prospectively from two devices: Zeiss PLEX Elite 9000 (Carl Zeiss Meditec, Dublin, CA) and Topcon DRI OCT Triton SS-OCTA (Topcon, Tokyo, Japan). Choroidal neovascularization area and VD of images were measured and analyzed with ImageJ software by two readers to evaluate the agreement between two devices, with respect to different image size (3 × 3 and 6 × 6 mm) and different image segmentation (automatic vs. manual), and a Topcon equivalent Zeiss segmentation as control (i.e., the equivalent anatomical slab given by Topcon device on the Zeiss device).

RESULTS

A total of 30 eyes (30 patients) were analyzed. There was an excellent agreement between the two readers in CNV area measurements intraclass correlation coefficient >0.9 in all analyses. We found excellent agreement in CNV area measurements (manual and automatic segmentations) when comparing 3 × 3-mm or 6 × 6-mm images both for each single device and between the two devices (overall intraclass correlation coefficient > 0.9). Vessel density measurements between manual to automatic segmentation within the same device and same image size had a high intraclass correlation coefficient value, but there was a poor agreement in VD between different image sizes (3 × 3 mm vs. 6 × 6 mm) in the same device and also comparing the two devices (3 × 3 Topcon vs. 3 × 3 Zeiss; 6 × 6 Topcon vs. 6 × 6 Zeiss). There was a poor agreement between the Topcon equivalent Zeiss segmentation and all other segmentations.

CONCLUSION

There was an excellent agreement in CNV area measurements for both swept-source optical coherence tomography angiography devices in automatic and manual segmentations. However, the Topcon equivalent Zeiss segmentation was not comparable with any of the preset segmentations of Topcon and Zeiss devices. There was a poor agreement in CNV VD between different image size and different devices. For these reasons, it seems that, for accurate longitudinal analysis of VD, it is better to use the same device for each individual, even if both devices can be used interchangeably for CNV area measurements using automatic or manual segmentations.

摘要

目的

比较两种不同的扫频源光相干断层扫描血管造影(SS-OCTA)设备的脉络膜新生血管(CNV)面积和血管密度(VD)测量值。

方法

前瞻性地从两台设备(德国卡尔蔡司公司的 Zeiss PLEX Elite 9000 和日本 Topcon 公司的 Topcon DRI OCT Triton SS-OCTA)收集受新生血管性年龄相关性黄斑变性影响的患者的面扫 OCTA 图像。两位读者使用 ImageJ 软件测量和分析图像的脉络膜新生血管面积和 VD,以评估两种设备之间的一致性,涉及不同的图像大小(3×3 和 6×6mm)和不同的图像分割(自动与手动),并以 Topcon 等效的 Zeiss 分割作为对照(即 Topcon 设备在 Zeiss 设备上提供的等效解剖切片)。

结果

共分析了 30 只眼(30 例患者)。在所有分析中,两位读者在 CNV 面积测量方面均具有极好的一致性(组内相关系数>0.9)。当比较每个设备的 3×3mm 或 6×6mm 图像以及比较两种设备时,我们发现手动和自动分割的 CNV 面积测量(整体组内相关系数>0.9)具有极好的一致性。同一设备内相同图像尺寸下手动到自动分割的 VD 测量值具有较高的组内相关系数值,但同一设备不同图像尺寸(3×3mm 与 6×6mm)之间以及比较两种设备(3×3Topcon 与 3×3Zeiss;6×6Topcon 与 6×6Zeiss)之间的 VD 存在较差的一致性。Topcon 等效的 Zeiss 分割与所有其他分割之间的一致性较差。

结论

两种 SS-OCTA 设备的自动和手动分割在 CNV 面积测量方面具有极好的一致性。然而,Topcon 等效的 Zeiss 分割与 Topcon 和 Zeiss 设备的任何预设分割均不具有可比性。不同图像尺寸和不同设备之间的 CNV VD 存在较差的一致性。因此,为了进行准确的 VD 纵向分析,对于每个个体而言,最好使用相同的设备,即使在使用自动或手动分割时,两种设备也可以互换用于 CNV 面积测量。