Abstract

BACKGROUND

Accurate and repeatable measurements of the retinal nerve fibre layer (RNFL) thickness are important in the diagnosis and management of glaucoma and other disorders.

OBJECTIVE

To assess the repeatability and reproducibility of the iVue-100 optical coherence tomographer (OCT).

METHODS

The thickness of the RNFL was measured for 50 healthy participants using the iVue-100 OCT. Although both eyes per participant were measured, only right eyes were analysed here. Repeatability and reproducibility of the iVue-100 OCT were assessed using the intraclass correlation coefficient (ICC), coefficient of variation (CoV), paired t-tests and Bland-Altman analysis.

RESULTS

Good intra-observer repeatability was obtained as indicated by the ICC of observer 1 (range: 0.941 - 0.976) and observer 2 (range: 0.829 - 0.953) as well by the CoV of observer 1 (range: 0.098 - 0.137) and observer 2 (0.091 - 0.132). In terms of inter-observer reproducibility, significant differences (p< 0.05) in mean measurements between the observers were noted for the average RNFL readings and in the superior and inferior quadrants as assessed with paired t-tests. Even though significant inter-session differences were found for the average RNFL thickness and the superior quadrant (p = 0.003 and p = 0.013, respectively), excellent ICCs were obtained for inter-session reproducibility (range: 0.914 - 0.979).

CONCLUSION

The iVue-100 OCT demonstrated good repeatability and reproducibility for RNFL thickness measurements.

摘要

背景

视网膜神经纤维层(RNFL)厚度的准确且可重复测量对于青光眼及其他疾病的诊断和管理至关重要。

目的

评估iVue - 100光学相干断层扫描仪(OCT)的重复性和再现性。

方法

使用iVue - 100 OCT对50名健康参与者的RNFL厚度进行测量。尽管对每位参与者的双眼都进行了测量,但此处仅分析右眼。使用组内相关系数(ICC)、变异系数(CoV)、配对t检验和布兰德 - 奥特曼分析来评估iVue - 100 OCT的重复性和再现性。

结果

观察者1的ICC(范围:0.941 - 0.976)和观察者2的ICC(范围:0.829 - 0.953)以及观察者1的CoV(范围:0.098 - 0.137)和观察者2的CoV(0.091 - 0.132)表明观察者内部具有良好的重复性。在观察者间再现性方面,通过配对t检验评估,观察者之间平均RNFL读数以及上下象限的平均测量值存在显著差异(p < 0.05)。尽管平均RNFL厚度和上象限在不同测量时段之间存在显著差异(分别为p = 0.003和p = 0.013),但不同测量时段的再现性获得了出色的ICC(范围:0.914 - 0.979)。

结论

iVue - 100 OCT在RNFL厚度测量方面显示出良好的重复性和再现性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/0347/6306963/86596ebabf01/AFHS1802-0304Fig1.jpg

相似文献

3
Repeatability of nerve fiber layer thickness measurements in patients with glaucoma and without glaucoma using spectral-domain and time-domain OCT.
Graefes Arch Clin Exp Ophthalmol. 2012 Feb;250(2):279-87. doi: 10.1007/s00417-011-1811-9. Epub 2011 Sep 10.
5
Comparison of retinal nerve fiber layer measurement between 2 spectral domain OCT instruments.
J Glaucoma. 2012 Apr-May;21(4):266-73. doi: 10.1097/IJG.0b013e3182071cdd.
6
Cirrus HD-OCT short-term repeatability of clinical retinal nerve fiber layer measurements.
Optom Vis Sci. 2015 Jan;92(1):83-8. doi: 10.1097/OPX.0000000000000452.
10
Optical coherence tomography of the retina, nerve fiber layer, and optic nerve head in dogs with glaucoma.
Vet Ophthalmol. 2020 Jan;23(1):97-112. doi: 10.1111/vop.12694. Epub 2019 Jul 12.

引用本文的文献

1
Characterization of Retinal Thickness in Individuals with Albinism: Baseline Data for a Black South African Population.
Clin Optom (Auckl). 2021 Jan 20;13:15-22. doi: 10.2147/OPTO.S273141. eCollection 2021.
2
Sexuality and other issues in Africa and beyond.
Afr Health Sci. 2018 Jun;18(2):i-iii. doi: 10.4314/ahs.v18i2.1.

本文引用的文献

1
Localized Changes in Retinal Nerve Fiber Layer Thickness as a Predictor of Localized Functional Change in Glaucoma.
Am J Ophthalmol. 2016 Oct;170:75-82. doi: 10.1016/j.ajo.2016.07.020. Epub 2016 Aug 1.
3
Understanding Bland Altman analysis.
Biochem Med (Zagreb). 2015 Jun 5;25(2):141-51. doi: 10.11613/BM.2015.015. eCollection 2015.
4
Correlation of retinal nerve fiber layer thickness and visual fields in glaucoma: a broken stick model.
Am J Ophthalmol. 2014 May;157(5):953-59. doi: 10.1016/j.ajo.2014.01.014. Epub 2014 Jan 30.
5
Retinal nerve fiber layer thickness, brain atrophy, and disability in multiple sclerosis patients.
J Neuroophthalmol. 2014 Mar;34(1):23-8. doi: 10.1097/WNO.0000000000000057.
6
Retinal nerve fiber layer thickness. The Beijing Eye Study 2011.
PLoS One. 2013 Jun 24;8(6):e66763. doi: 10.1371/journal.pone.0066763. Print 2013.
7
The molecular basis of retinal ganglion cell death in glaucoma.
Prog Retin Eye Res. 2012 Mar;31(2):152-81. doi: 10.1016/j.preteyeres.2011.11.002. Epub 2011 Dec 4.
8
Global estimates of visual impairment: 2010.
Br J Ophthalmol. 2012 May;96(5):614-8. doi: 10.1136/bjophthalmol-2011-300539. Epub 2011 Dec 1.
9
Intra and interoperator reproducibility of retinal nerve fibre and macular thickness measurements using Cirrus Fourier-domain OCT.
Acta Ophthalmol. 2011 Feb;89(1):e23-9. doi: 10.1111/j.1755-3768.2010.02045.x. Epub 2010 Nov 25.
10
Reproducibility of retinal thickness measurements on normal and pathologic eyes by different optical coherence tomography instruments.
Am J Ophthalmol. 2010 Dec;150(6):815-24. doi: 10.1016/j.ajo.2010.06.025. Epub 2010 Oct 20.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验