Accurate and repeatable measurements of the retinal nerve fibre layer (RNFL) thickness are important in the diagnosis and management of glaucoma and other disorders.
To assess the repeatability and reproducibility of the iVue-100 optical coherence tomographer (OCT).
The thickness of the RNFL was measured for 50 healthy participants using the iVue-100 OCT. Although both eyes per participant were measured, only right eyes were analysed here. Repeatability and reproducibility of the iVue-100 OCT were assessed using the intraclass correlation coefficient (ICC), coefficient of variation (CoV), paired t-tests and Bland-Altman analysis.
Good intra-observer repeatability was obtained as indicated by the ICC of observer 1 (range: 0.941 - 0.976) and observer 2 (range: 0.829 - 0.953) as well by the CoV of observer 1 (range: 0.098 - 0.137) and observer 2 (0.091 - 0.132). In terms of inter-observer reproducibility, significant differences (p< 0.05) in mean measurements between the observers were noted for the average RNFL readings and in the superior and inferior quadrants as assessed with paired t-tests. Even though significant inter-session differences were found for the average RNFL thickness and the superior quadrant (p = 0.003 and p = 0.013, respectively), excellent ICCs were obtained for inter-session reproducibility (range: 0.914 - 0.979).
The iVue-100 OCT demonstrated good repeatability and reproducibility for RNFL thickness measurements.
视网膜神经纤维层(RNFL)厚度的准确且可重复测量对于青光眼及其他疾病的诊断和管理至关重要。
评估iVue - 100光学相干断层扫描仪(OCT)的重复性和再现性。
使用iVue - 100 OCT对50名健康参与者的RNFL厚度进行测量。尽管对每位参与者的双眼都进行了测量,但此处仅分析右眼。使用组内相关系数(ICC)、变异系数(CoV)、配对t检验和布兰德 - 奥特曼分析来评估iVue - 100 OCT的重复性和再现性。
观察者1的ICC(范围:0.941 - 0.976)和观察者2的ICC(范围:0.829 - 0.953)以及观察者1的CoV(范围:0.098 - 0.137)和观察者2的CoV(0.091 - 0.132)表明观察者内部具有良好的重复性。在观察者间再现性方面,通过配对t检验评估,观察者之间平均RNFL读数以及上下象限的平均测量值存在显著差异(p < 0.05)。尽管平均RNFL厚度和上象限在不同测量时段之间存在显著差异(分别为p = 0.003和p = 0.013),但不同测量时段的再现性获得了出色的ICC(范围:0.914 - 0.979)。
iVue - 100 OCT在RNFL厚度测量方面显示出良好的重复性和再现性。