Helfen Anne, Masthoff Max, Claussen Jing, Gerwing Mirjam, Heindel Walter, Ntziachristos Vasilis, Eisenblätter Michel, Köhler Michael, Wildgruber Moritz
Institute of Clinical Radiology, Medical Faculty, University of Muenster, and University Hospital Muenster, 48149 Muenster, Germany.
iThera Medical, 81379 Munich, Germany.
J Clin Med. 2019 Jan 9;8(1):63. doi: 10.3390/jcm8010063.
Multispectral optoacoustic tomography (MSOT) represents a new imaging approach revealing functional tissue information without extrinsic contrast agents. Using a clinical combined ultrasound (US)/MSOT device, we investigated the interindividual robustness and impact of intra- and interobserver variability of MSOT values in soft tissue (muscle and subcutaneous fat) of healthy volunteers. Semiquantitative MSOT values for deoxygenated (Hb), oxygenated (HbO₂) and total hemoglobin (HbT), as well as oxygen saturation (sO₂), were calculated for both forearms in transversal and longitudinal probe orientation ( = 3, 8 measurements per subject). For intraobserver reproducibility, the same examiner investigated three subjects twice. Mean values of left vs. right forearm and transversal vs. longitudinal probe orientation were compared using an unpaired Student's test. Bland Altmann plots with 95% limits of agreement for absolute averages and differences were calculated. Intraclass correlation coefficients (ICC 2,k) were computed for three different examiners. We obtained reproducible and consistent MSOT values with small-to-moderate deviation for muscle and subcutaneous fat tissue. Probe orientation and body side had no impact on calculated MSOT values ( > 0.05 each). Intraobserver reproducibility revealed equable mean values with small-to-moderate deviation. For muscular tissue, good ICC was obtained for sO₂. Measurements of subcutaneous tissue revealed good-to-excellent ICCs for all calculated values. Thus, in this preliminary study on healthy individuals, clinical MSOT provided consistent and reproducible functional soft tissue characterization, independent on the investigating personnel.
多光谱光声断层扫描(MSOT)是一种无需外源性造影剂就能揭示组织功能信息的新型成像方法。我们使用临床超声(US)/MSOT联合设备,研究了健康志愿者软组织(肌肉和皮下脂肪)中MSOT值的个体间稳健性以及观察者内和观察者间变异性的影响。在横向和纵向探头方向上,对两名受试者的双侧前臂计算了脱氧血红蛋白(Hb)、氧合血红蛋白(HbO₂)和总血红蛋白(HbT)的半定量MSOT值以及氧饱和度(sO₂)(每名受试者 = 3次,每次测量8次)。为了评估观察者内的可重复性,同一名检查者对三名受试者进行了两次检查。使用不成对的Student's t检验比较左右前臂以及横向与纵向探头方向的平均值。计算了绝对平均值和差异的一致性界限为95%的Bland Altmann图。计算了三名不同检查者的组内相关系数(ICC 2,k)。我们获得了可重复且一致的MSOT值,肌肉和皮下脂肪组织的偏差较小至中等。探头方向和身体部位对计算出的MSOT值没有影响(每次P > 0.05)。观察者内的可重复性显示平均值均匀,偏差较小至中等。对于肌肉组织,sO₂的ICC良好。皮下组织测量显示所有计算值的ICC均为良好至优秀。因此,在这项针对健康个体的初步研究中,临床MSOT提供了一致且可重复的功能性软组织特征描述,与研究人员无关。