Kim Kang Sup, Bae Woong Jin, Kim Sae Woong, Lee Moo Yeon
Department of Urology, Incheon St. Mary's Hospital, College of Medicine, The Catholic University of Korea, 59, Dongsu-ro, Bupyeong-gu, Incheon, 21431, Republic of Korea.
Department of Urology, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, 222, Banpo-daero, Seocho-gu, Seoul, 06591, Republic of Korea.
BMC Urol. 2019 Jan 16;19(1):6. doi: 10.1186/s12894-018-0425-5.
The purpose of this study was to report the outcomes of patients who underwent penile implantation with AMS 700 LGX inflatable penile prosthesis (IPP) in a single center by a single surgeon.
A total of 342 patients with erectile dysfunction who underwent AMS 700 LGX IPP implantation between October 2014 and April 2016 were included in this study. All patients were evaluated using the International Index of Erectile Function questionnaire preoperatively and at 3, 6, and 12 months postoperatively. We also investigated the mean stretched flaccid penile length before and after surgery as well as the complications related to and the mechanical reliability of the IPP.
The questionnaire scores at 12 months were statistically significantly higher than the baseline scores. The mean stretched flaccid penile length was 11.1 ± 0.8 cm at baseline and was longer at 3 (11.9 ± 0.9 cm, P < 0.001), 6 (12.0 ± 0.9 cm, P < 0.001), and 12 (12.2 ± 0.7 cm, P < 0.001) postoperatively. There were no intraoperative or perioperative complications. However, one patient had infection and 10 patients developed mechanical failure during the follow-up duration.
The results of our study suggest that the AMS 700 LGX IPP could be used to prevent penile shortening in patients undergoing IPP implantation. Furthermore, erectile function and patient satisfaction were improved excellently.
本研究旨在报告在单一中心由单一外科医生为患者植入美国医疗系统公司(AMS)700 LGX可膨胀阴茎假体(IPP)的治疗结果。
本研究纳入了2014年10月至2016年4月期间接受AMS 700 LGX IPP植入的342例勃起功能障碍患者。所有患者在术前以及术后3、6和12个月使用国际勃起功能指数问卷进行评估。我们还调查了手术前后阴茎疲软拉伸平均长度以及与IPP相关的并发症和机械可靠性。
术后12个月的问卷评分在统计学上显著高于基线评分。阴茎疲软拉伸平均长度在基线时为11.1±0.8厘米,术后3个月(11.9±0.9厘米,P<0.001)、6个月(12.0±0.9厘米,P<0.001)和12个月(12.2±0.7厘米,P<0.001)时均更长。术中或围手术期无并发症。然而,在随访期间有1例患者发生感染,10例患者出现机械故障。
我们的研究结果表明,AMS 700 LGX IPP可用于防止接受IPP植入的患者阴茎缩短。此外,勃起功能和患者满意度得到了显著改善。