电催化葡萄糖传感器。

The electrocatalytic glucose sensor.

作者信息

von Lucadou I, Luft G, Preidel W, Richter G J

机构信息

Siemens Research and Development Laboratories, Erlangen, Germany.

出版信息

Horm Metab Res Suppl. 1988;20:41-3.

PMID:3074037
Abstract

An electrocatalytic glucose sensor is a very useful component of an artificial beta cell in controlling the insulin dosage in the diabetes therapy. The advantage is its potential long-term stability and the disadvantage is its lack of selectivity. The principle and experimental results with different electrolytes will be described. In saline solution the mean deviation in glucose response is up to 15% if the amino acid and urea concentrations raise from minimum to their physiological maximum values.

摘要

电催化葡萄糖传感器是人工β细胞在糖尿病治疗中控制胰岛素剂量的非常有用的组件。其优点是具有潜在的长期稳定性,缺点是缺乏选择性。将描述其原理以及使用不同电解质的实验结果。在盐溶液中,如果氨基酸和尿素浓度从最小值升至其生理最大值,则葡萄糖响应的平均偏差高达15%。