The equilibration time of an autoclave is defined as the period of time that elapses between the attainment of the sterilization temperature at the reference measurement point of the autoclave chamber and the attainment of the sterilization temperature at all measurement probes within the dry goods load placed within the autoclave during a sterilization cycle.The equilibration time is an indicator of the ability of the autoclave to remove air from and to subsequently sterilize the dry goods load placed within the autoclave chamber. Therefore, the equilibration time is an indicator of the performance of an autoclave. Periodic measurements of the equilibration time and evaluation of the equilibration time data can assure continued performance of the autoclave.For autoclaves that support biopharmaceutical manufacturing plants, the equilibration time is generally measured by placing heat penetration probes within the dry goods load items placed within the autoclave during a sterilization cycle. However, measurement of the equilibration time in this fashion yields results that are highly variable and difficult to replicate.This paper presents a study that attempts to understand the impact of the following autoclave variables on the consistent measurement of the equilibration time and on the reduction of the equilibration time: type of dry goods load item placed inside the autoclave, location of the heat penetration probe placed within the dry goods load item, location of the dry goods load item inside the autoclave chamber, orientation of the dry goods load item inside the autoclave chamber, number of coils of the tubing elements of a dry goods load item, total thermal mass of the dry goods load items placed inside the autoclave, temperature of the dry goods load items at the start of the sterilization cycle, presence of residual moisture in the dry goods load items at the start of the sterilization cycle, number of steam and vacuum pulses used for preconditioning in the autoclave recipe, depth of the steam and vacuum pulses used for preconditioning in the autoclave recipe, and hold time of the steam and vacuum pulses used for preconditioning in the autoclave recipe.The focus of this paper is on tests performed to evaluate these variables followed by analysis and interpretation of the test data to identify key variables that have the highest impact on the measurement of the equilibration time. The conclusion of this paper presents a process to achieve shorter equilibration times in autoclaves through control of the key variables identified. The equilibration time is an indicator of the performance of an autoclave. Periodic measurements of the equilibration time and evaluation of the equilibration time data of an autoclave can assure continued performance of said autoclave.This paper presents a study performed to develop a general guideline for consistent measurement of the equilibration time of an autoclave through identification and control of key sources of variability that impact the measurement. The focus of this paper is the testing and data analysis performed to identify these key sources of variability observed in a typical autoclave load sterilized through a typical autoclave cycle in a biopharmaceutical manufacturing plant. The conclusion of this paper presents a process to achieve shorter equilibration times in autoclaves through control of the key variables identified.

摘要

高压灭菌器的平衡时间定义为

在灭菌循环过程中,从高压灭菌器腔室的参考测量点达到灭菌温度起,到放置在高压灭菌器内的干货负载中的所有测量探头达到灭菌温度所经过的时间段。平衡时间是高压灭菌器从其腔室内去除空气并随后对放置其中的干货负载进行灭菌能力的一个指标。因此,平衡时间是高压灭菌器性能的一个指标。定期测量平衡时间并评估平衡时间数据可以确保高压灭菌器持续保持性能。对于支持生物制药生产厂的高压灭菌器,平衡时间通常是通过在灭菌循环期间将热穿透探头放置在置于高压灭菌器内的干货负载物品中来测量的。然而,以这种方式测量平衡时间会产生高度可变且难以重复的结果。本文介绍了一项研究,该研究试图了解以下高压灭菌器变量对平衡时间的一致测量以及平衡时间缩短的影响:放置在高压灭菌器内的干货负载物品的类型、放置在干货负载物品内的热穿透探头的位置、干货负载物品在高压灭菌器腔室内的位置、干货负载物品在高压灭菌器腔室内的方向、干货负载物品的管件的盘管数量、放置在高压灭菌器内的干货负载物品的总热质量、灭菌循环开始时干货负载物品的温度、灭菌循环开始时干货负载物品中残留水分的存在情况、高压灭菌器配方中用于预处理的蒸汽和真空脉冲的数量、高压灭菌器配方中用于预处理的蒸汽和真空脉冲的深度,以及高压灭菌器配方中用于预处理的蒸汽和真空脉冲的保持时间。本文的重点是进行测试以评估这些变量,随后对测试数据进行分析和解释,以确定对平衡时间测量影响最大的关键变量。本文的结论提出了一个通过控制已确定的关键变量来在高压灭菌器中实现更短平衡时间的过程。平衡时间是高压灭菌器性能的一个指标。定期测量高压灭菌器的平衡时间并评估其平衡时间数据可以确保该高压灭菌器持续保持性能。本文介绍了一项为制定通过识别和控制影响测量的关键变异性来源来一致测量高压灭菌器平衡时间的通用指南而进行的研究。本文的重点是在生物制药生产厂中通过典型的高压灭菌器循环对典型的高压灭菌器负载进行灭菌时,为识别观察到的这些关键变异性来源而进行的测试和数据分析。本文的结论提出了一个通过控制已确定的关键变量来在高压灭菌器中实现更短平衡时间的过程。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验