Yilmaz Oznur, Karapinar Tolga
Department of Internal Medicine, Faculty of Veterinary Medicine, Firat University, Elazig, Turkey.
Vet Clin Pathol. 2019 Mar;48(1):31-35. doi: 10.1111/vcp.12705. Epub 2019 Feb 20.
Information regarding the performance of the i-STAT, a point-of-care analyzer, to determine ionized calcium (iCa ) concentration in bovine blood is limited.
The primary goal of this study was to evaluate the agreement of the i-STAT against a traditional bench-top blood gas analyzer (Radiometer ABL800 FLEX) in the measurement of iCa in the blood of cattle with various disorders.
In this study, 121 cattle with various diseases were used. Venous blood samples were run on the i-STAT and the Radiometer ABL800 FLEX for determination of blood iCa concentrations. Data were subjected to Passing-Bablok regression and Bland-Altman plots. The Radiometer ABL800 FLEX was the reference method and the i-STAT was the test method. The i-STAT performance was evaluated at two different cutoff values for iCa concentrations (<1.00 and <1.18 mmol/L) using the receiver operating characteristic curve.
The Passing-Bablok regression equation that fit the iCa concentration as measured with the i-STAT (Y) and the Radiometer ABL800 FLEX (X), was Y = -0.108276 [-0.15/-0.06] + 1.103448 [1.06/1.14] × X, with a residual deviation of 0.02. The deviation from the identity line was not significant (P = 0.64). The Bland-Altman difference plot revealed no bias between the i-STAT and the Radiometer ABL 800 FLEX. The i-STAT had a sensitivity of 100% (95% CI, 87.7-100.0) and 94.8% (95% CI, 87.2-98.6) at the iCa concentration cutoff values of <1.00 and <1.18 mmol/L, respectively.
iCa concentrations in bovine blood measured with the i-STAT agreed with those of the Radiometer ABL800 FLEX. In particular, the i-STAT predicted hypocalcemia with a high sensitivity.
关于即时检验分析仪i-STAT测定牛血液中离子钙(iCa)浓度的性能信息有限。
本研究的主要目的是评估i-STAT与传统台式血气分析仪(Radiometer ABL800 FLEX)在测量患有各种疾病的牛血液中iCa时的一致性。
本研究使用了121头患有各种疾病的牛。采集静脉血样,分别用i-STAT和Radiometer ABL800 FLEX测定血液iCa浓度。数据进行了Passing-Bablok回归分析和Bland-Altman图分析。以Radiometer ABL800 FLEX为参考方法,i-STAT为测试方法。使用受试者工作特征曲线在两个不同的iCa浓度临界值(<1.00和<1.18 mmol/L)下评估i-STAT的性能。
i-STAT测量的iCa浓度(Y)与Radiometer ABL800 FLEX测量的iCa浓度(X)的Passing-Bablok回归方程为Y = -0.108276 [-0.15/-0.06] + 1.103448 [1.06/1.14]×X,残余偏差为0.02。与恒等线的偏差不显著(P = 0.64)。Bland-Altman差异图显示i-STAT与Radiometer ABL 800 FLEX之间无偏差。在iCa浓度临界值<1.00和<1.18 mmol/L时,i-STAT的灵敏度分别为100%(95%CI,87.7 - 100.0)和94.8%(95%CI,87.2 - 98.6)。
用i-STAT测量的牛血液中iCa浓度与Radiometer ABL800 FLEX测量的结果一致。特别是,i-STAT预测低钙血症的灵敏度较高。