评估4种即时检测设备用于测定牛血液中l-乳酸浓度的情况。

Evaluation of 4 point-of-care units for the determination of blood l-lactate concentration in cattle.

作者信息

Karapinar T, Kaynar O, Hayirli A, Kom M

机构信息

Faculty of Veterinary Medicine, Department of Internal Medicine, Firat University, Elazig, Turkey.

出版信息

J Vet Intern Med. 2013 Nov-Dec;27(6):1596-603. doi: 10.1111/jvim.12204. Epub 2013 Oct 1.

Abstract

BACKGROUND

Despite being used commonly in bovine medicine, information on reliability of point-of-care (POC) lactate meters is limited.

OBJECTIVE

To determine the validity of 4 commercially available POC lactate meters in cattle.

ANIMALS

Cattle with various diseases (n = 120).

METHODS

Blood samples collected from the jugular vein were processed immediately on POC lactate meters. Plasma l-lactate concentration was measured by the enzymatic-colorimetric method (ELISA). Data were subjected to Friedman's test for comparison, Passing-Bablok regression and Bland-Altman plot analyses for reliability, and receiver operating characteristics analysis for sensitivity (Se) and specificity (Sp).

RESULTS

The POC lactate meters were highly correlated with ELISA (r = 0.98-0.99) despite disagreements among units. When regressed on ELISA, blood l-lactate concentrations generated from Accutrend Plus and Lactate Pro were linear up to 16.6 and 15.7 mmol/L, respectively, whereas those generated from i-STAT and Lactate Scout were linear up to 19.5 and 19.7 mmol/L, respectively. All POC lactate meters had a Se of 100% with Sp of 95.7-98.6% at a plasma l-lactate cut-off concentration of 4 mmol/L. i-STAT had the best accuracy (99.0%) and precision (99.8%), the best linear fit (y = -0.13 + 1.04X) yielding the lowest bias (-6.6%) as well as the highest Se (100%) and Sp (98.6%).

CONCLUSIONS AND CLINICAL IMPORTANCE

Despite high correlation with the reference method, dilution is needed for Accutrend Plus/Lactate Pro and i-STAT/Lactate Scout if concentrations >15 and 20 mmol/L, respectively. i-STAT provided the most accurate and precise results.

摘要

背景

即时检测(POC)乳酸测定仪虽在牛医学中广泛应用,但其可靠性信息有限。

目的

确定4种市售POC乳酸测定仪在牛身上的有效性。

动物

患有各种疾病的牛(n = 120)。

方法

从颈静脉采集的血样立即用POC乳酸测定仪进行检测。采用酶比色法(ELISA)测定血浆L-乳酸浓度。数据进行Friedman检验以比较,Passing-Bablok回归和Bland-Altman图分析以评估可靠性,以及受试者工作特征分析以评估敏感性(Se)和特异性(Sp)。

结果

尽管各仪器之间存在差异,但POC乳酸测定仪与ELISA高度相关(r = 0.98 - 0.99)。以ELISA为参照进行回归分析时,Accutrend Plus和Lactate Pro产生的血L-乳酸浓度分别在16.6和15.7 mmol/L以内呈线性,而i-STAT和Lactate Scout产生的血L-乳酸浓度分别在19.5和19.7 mmol/L以内呈线性。在血浆L-乳酸截断浓度为4 mmol/L时,所有POC乳酸测定仪的敏感性均为100%,特异性为95.7% - 98.6%。i-STAT具有最佳的准确性(99.0%)和精密度(99.8%),最佳线性拟合(y = -0.13 + 1.04X),偏差最低(-6.6%),敏感性(100%)和特异性(9%)最高。

结论及临床意义

尽管与参考方法高度相关,但如果浓度分别>15和20 mmol/L,Accutrend Plus/Lactate Pro和i-STAT/Lactate Scout需要进行稀释。i-STAT提供了最准确和精确的结果。