Elworthy P M
Neurochem Res. 1986 May;11(5):753-8. doi: 10.1007/BF00965343.
Catecholamines extracted from tissues are readily measured by fluorescence following trihydroxyindole derivatization. A complicating factor, however, has been the variably high background fluorescence levels obtained which adversely affect the precision of the assay. Employing EDTA is shown to reduce the high background fluorescence to a consistently low level, and to improve the accuracy of the trihydroxyindole method. It is suggested that high background fluorescence readings obtained by previous workers were due to metal ions extracted from tissue and carried over into the derivatization procedure.
从组织中提取的儿茶酚胺在经三羟基吲哚衍生化后,很容易通过荧光法进行测定。然而,一个复杂的因素是所获得的背景荧光水平总是很高,这对测定的精度产生了不利影响。结果表明,使用乙二胺四乙酸(EDTA)可将高背景荧光降低至始终较低的水平,并提高三羟基吲哚法的准确性。有人认为,以前的研究人员获得的高背景荧光读数是由于从组织中提取并带入衍生化过程的金属离子所致。