Pretreatment Dose Verification in Volumetric Modulated Arc Therapy Using Liquid Ionization Chamber.

作者信息

Kakade Nitin R, Kumar Rajesh, Sharma Sunil Dutt, Mittal Vikram, Datta D

机构信息

Radiological Physics and Advisory Division, Bhabha Atomic Research Center, Anushaktinagar, Mumbai, Maharashtra, India.

Homi Bhabha National Institute, Anushaktinagar, Mumbai, Maharashtra, India.

出版信息

J Med Phys. 2019 Jan-Mar;44(1):9-15. doi: 10.4103/jmp.JMP_108_18.

Abstract

PURPOSE

The purpose of the present study was to evaluate the practicability of liquid ionization chamber (LIC) for pretreatment dose verification of the advanced radiotherapy techniques such as volumetric modulated arc therapy (VMAT).

MATERIALS AND METHODS

The dosimetric characteristics of LIC such as repeatability, sensitivity, monitor unit linearity, dose rate dependence, angular dependence, voltage-current response, and output factors were investigated in 6 MV therapeutic X-ray beams. The LIC was cross-calibrated against 0.125-cc air-filled thimble ionization chamber. A dedicated dosimetry insert made up of Perspex to incorporate the LIC at proper location in the intensity-modulated radiation therapy thorax phantom was locally fabricated. The collection efficiency and ion recombination correction factor was determined using the two-dose rate method. Pretreatment dose verification measurement of VMAT treatment plans were carried out using the liquid ionization chamber as well as small volume (0.125 cc) air-filled thimble ionization chamber. The measured dose values by the two dosimeters and TPS calculated dose at a given point were compared.

RESULTS

The relative percentage differences between the TPS calculated and measured doses were within ± 1.57% for LIC and ± 2.21% for 0.125 cc ionization chamber, respectively.

CONCLUSIONS

The measured dose values by the two dosimeters and TPS calculated dose at a given point were found comparable suggesting that the LIC could be a good choice of dosimeter for pretreatment dose verification in VMAT.

摘要

目的

本研究的目的是评估液体电离室(LIC)在容积调强弧形放疗(VMAT)等先进放疗技术预处理剂量验证中的实用性。

材料与方法

在6MV治疗性X射线束中研究了LIC的剂量学特性,如重复性、灵敏度、监测单位线性、剂量率依赖性、角度依赖性、电压-电流响应和输出因子。LIC与0.125cc空气填充指形电离室进行交叉校准。局部制作了一个由有机玻璃制成的专用剂量学插入件,以便将LIC放置在调强放射治疗胸部体模的适当位置。使用双剂量率法确定收集效率和离子复合校正因子。使用液体电离室以及小体积(0.125cc)空气填充指形电离室对VMAT治疗计划进行预处理剂量验证测量。比较了两种剂量计测量的剂量值与TPS在给定点计算的剂量。

结果

LIC的TPS计算剂量与测量剂量之间的相对百分比差异分别在±1.57%以内,0.125cc电离室在±2.21%以内。

结论

发现两种剂量计测量的剂量值与TPS在给定点计算的剂量具有可比性,这表明LIC可能是VMAT预处理剂量验证的良好剂量计选择。