Lucas A, Nohria V, Roberts S B
Biol Neonate. 1987;51(3):138-43. doi: 10.1159/000242644.
A new method is described for measuring the rate of carbon dioxide production, and hence for estimating energy expenditure, in preterm infants receiving assisted ventilation. In a validation study, the mean error in carbon dioxide measurement was 1.9%. Measurements were made, over a 45-min period, on 11 sick, ventilated subjects and carbon dioxide production rate was 5.2 +/- 0.7 (SD) ml/min X kg body weight. We suggest that continuous monitoring of carbon dioxide output will contribute to the clinical assessment of the effects of different ventilator settings on pulmonary gas exchange and that estimated values for energy expenditure will be of value in nutritional studies on sick ventilated infants.