Hamilton R D, Crockett R J, Alpers J H
Anaesth Intensive Care. 1978 Aug;6(3):251-5. doi: 10.1177/0310057X7800600314.
The effect of sodium heparin on the determination of the major arterial blood gas parameters was studied. The addition of heparin produced errors in all three parameters tested, i.e. PO2, PCO2 and pH. The PCO2 determination was most affected, a 12% decrease being introduced into a 1.5 ml blood sample by the volume of heparin which fills the deadspace of a syringe and needle. Small but statistically significant errors were also produced in PO2 and pH by this heparin concentration. The clinical implications of these findings are discussed and certain measures outlined to avoid such errors.
研究了肝素钠对主要动脉血气参数测定的影响。肝素的添加在所有三个测试参数(即PO2、PCO2和pH)中都产生了误差。PCO2测定受影响最大,填充注射器和针头死腔的肝素体积会使1.5 ml血样中的PCO2降低12%。这种肝素浓度在PO2和pH中也产生了虽小但具有统计学意义的误差。讨论了这些发现的临床意义,并概述了某些避免此类误差的措施。