Suppr超能文献

Serum exchange and use of dilutions have improved precision of measurement of islet cell antibodies.

作者信息

Bonifacio E, Lernmark A, Dawkins R L

机构信息

Department of Clinical Immunology, Royal Perth Hospital, Western Australia.

出版信息

J Immunol Methods. 1988 Jan 21;106(1):83-8. doi: 10.1016/0022-1759(88)90274-8.

Abstract

In an attempt to improve the diagnostic value of measuring antibodies to islet cell cytoplasmic antigen, coded sera were distributed to 38 laboratories and results were returned for analysis. Comparison between laboratories revealed that results for some individual sera differed by up to nine doubling dilutions and even within laboratories duplicate samples could differ by six doubling dilutions. By including dilutions of sera it was possible to draw a standard curve for each laboratory and this revealed major variations in shape, slope and intercept. However, using each laboratory's standard curve and converting results to units, a substantial improvement was obtained. The approach described improves standardisation and will permit laboratories to identify poor precision and/or any systematic change in assay performance.