Roy Jean-Philippe, Reagor James, Klein Tom, Morgan Jolyn, Palazzolo Alisa, Ratcliff Joseph, Goldstein Stuart L
Center for Acute Care Nephrology, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA.
Department of Cardiovascular Perfusion, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA.
Ther Apher Dial. 2020 Apr;24(2):197-201. doi: 10.1111/1744-9987.13349. Epub 2019 Jul 29.
Children on extracorporeal devices are at increased risk of hypothermia and require circuits with lower extracorporeal volume to avoid blood priming. We performed an in vitro study of the enFlow blood warmer to assess its warming performance and impact on circuit resistance at a high blood flow rate. The enFlow was added on the return line of a continuous renal replacement therapy circuit in a closed circuit primed with expired packed red blood cells (40% Hct). Return venous pressure and temperature pre- and post- both enFlow and Prismaflo II were measured at varying blood flow with different combination of the two blood warmers. Each variable was recorded eight times over 2 min. Return venous pressure was higher with the enFlow but was still within clinically usable range (127.5 ± 4.6 vs. 45.3 ± 1.4 mm Hg at 200 mL/min, P < 0.05). The enFlow was able to achieve post-warmer temperature between 34.1 and 35.2°C even at 200 mL/min with blood, lower than what was observed with crystalloid, but still more effective than the Prismaflo II (P < 0.05 vs. enFlow post-warmer temperature) achieving between 32.8 and 34.0°C. Combining both warmers achieved a higher post-warmer temperature between 35.1 and 36.5°C. The enFlow can warm blood at high flow rates with minimal extracorporeal volume increase (~5 mL) and improve hypothermia prevention but may have resistance issues at the highest flows. The use of this device could benefit pediatric practice where smaller extracorporeal volumes are needed and patients are at increased risk of hypothermia.
使用体外设备的儿童体温过低风险增加,需要体外循环血量较低的回路以避免血液预充。我们对enFlow血液加温器进行了一项体外研究,以评估其在高血流速率下的加温性能及其对回路阻力的影响。enFlow被添加到一个用过期浓缩红细胞(血细胞比容40%)预充的连续肾脏替代治疗回路的回流管线上。在不同血流速率下,使用两种血液加温器的不同组合,测量了enFlow和Prismaflo II前后的回静脉压和温度。每个变量在2分钟内记录8次。使用enFlow时回静脉压较高,但仍在临床可用范围内(200毫升/分钟时为127.5±4.6毫米汞柱,而Prismaflo II为45.3±1.4毫米汞柱,P<0.05)。即使在血流速度为200毫升/分钟时,enFlow也能够使加温器后的温度达到34.1至35.2°C,低于使用晶体液时观察到的温度,但仍比Prismaflo II更有效(与enFlow加温器后温度相比,P<0.05),后者达到32.8至34.0°C。两种加温器联合使用可使加温器后的温度更高,达到35.1至36.5°C。enFlow能够在高血流速率下加温血液,体外循环血量增加最小(约5毫升),并改善体温过低的预防,但在最高流速时可能存在阻力问题。该设备的使用可能有益于儿科实践,因为儿科需要较小的体外循环血量,且患者体温过低风险增加。