在恶劣条件下,比较 Buddy Lite、enFlow 和 Thermal Angel 对美军医疗人员的升温能力:文献综述。

Comparison of Warming Capabilities Between Buddy Lite, enFlow, and Thermal Angel for US Army Medical Personnel in Austere Conditions: A Literature Review.

出版信息

J Spec Oper Med. 2022 Dec 16;22(4):9-13. doi: 10.55460/HFTS-LQNF.

Abstract

US Army Forward Surgical Elements (FSEs) are highly mobile teams that provide damage control surgery (DCS) and damage control resuscitation (DCR) in austere locations that often lack standard hospital utilities (electricity, heat, food, and water). FSEs rely on portable battery-operated intravenous (IV) fluid warmers to remain light and mobile. However, their ability to warm blood in a massive resuscitation requires additional analysis. The purpose of this literature review is to examine the three most common battery-operated IV fluid warmers as determined by type and quantity listed on the Mission Table of Organization and Equipment (MTOE) of organic mobile medical units. These include the Buddy Lite, enFlow, and Thermal Angel, which are available to deployed US Army FSEs for blood resuscitation therapy. Based on limited available evidence, the enFlow produced higher outlet temperatures, effectively warmed greater volumes, reached the time to peak temperature faster, and produced greatest flow rates, with cool saline (5-10°C), compared to the Thermal Angel and Buddy Lite. However, recently the US Food and Drug Administration (FDA) issued a Class 1 recall on enFlow cartridges. Testing demonstrated aluminum elution from enFlow cartridges into IV solutions, thereby exposing patients to potentially unsafe aluminum levels. The authors recommend FSE units conduct a 100% enFlow cartridge inventory and seek an alternative IV fluid warming system prior to enFlow cartridge disposal. If an alternative does not exist, or the alternative warming system does not fit mission requirements, then medical personnel must carefully weigh the risks and benefits associated with the enFlow delivery system.

摘要

美国陆军前方外科手术单元 (FSE) 是高度机动的团队,在缺乏标准医院设施(电力、热力、食物和水)的恶劣环境中提供损伤控制外科手术 (DCS) 和损伤控制复苏 (DCR)。FSE 依靠便携式电池供电的静脉 (IV) 液体加热器来保持轻便和机动性。然而,它们在大规模复苏中加热血液的能力需要进一步分析。本文献综述的目的是检查根据有机机动医疗单位的组织和设备任务表 (MTOE) 列出的类型和数量确定的三种最常见的电池供电 IV 液体加热器。这些包括 Buddy Lite、enFlow 和 Thermal Angel,它们可供部署的美国陆军 FSE 用于血液复苏治疗。基于有限的可用证据,enFlow 产生了更高的出口温度,有效地加热了更大的体积,更快地达到了峰值温度,并且产生了最大的流速,与 Thermal Angel 和 Buddy Lite 相比,使用冷却的生理盐水(5-10°C)。然而,最近美国食品和药物管理局 (FDA) 对 enFlow 药筒发布了一级召回。测试表明 enFlow 药筒中的铝洗脱到 IV 溶液中,从而使患者面临潜在的不安全铝水平。作者建议 FSE 单位对 enFlow 药筒进行 100%的库存盘点,并在处理 enFlow 药筒之前寻找替代的 IV 液体加热系统。如果不存在替代方案,或者替代加热系统不符合任务要求,那么医务人员必须仔细权衡与 enFlow 输送系统相关的风险和收益。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验