Department of Medical Affairs, Vyaire Medical, 26125 North Riverwoods Blvd, Mettawa, IL, 60045, USA.
BMC Anesthesiol. 2022 Feb 10;22(1):44. doi: 10.1186/s12871-022-01585-w.
Perioperative hypothermia is a common occurrence, particularly with the elderly and pediatric age groups. Hypothermia is associated with an increased risk of perioperative complications. One method of preventing hypothermia is warming the infused fluids given during surgery. The enFlow™ intravenous fluid warmer has recently been reintroduced with a parylene coating on its heating blocks. In this paper, we evaluated the impact of the parylene coating on the new enFlow's fluid warming capacity.
Six coated and six uncoated enFlow cartridges were used. A solution of 10% propylene glycol and 90% distilled HO was infused into each heating cartridge at flow rates of 2, 10, 50, 150, and 200 ml/min. The infused fluid temperature was set at 4 °C, 20 °C, and 37 °C. Output temperature was recorded at each level. Data for analysis was derived from 18 runs at each flow rate (six cartridges at three temperatures).
The parylene coated fluid warming cartridge delivered very stable output of 40 °C temperatures at flow rates of 2, 10, and 50 ml/min regardless of the temperature of the infusate. At higher flow rates, the cartridges were not able to achieve the target temperature with the colder fluid. Both cartridges performed with similar efficacy across all flow rates at all temperatures.
At low flow rates, the parylene coated enFlow cartridges was comparable to the original uncoated cartridges. At higher flow rates, the coated and uncoated cartridges were not able to achieve the target temperature. The parylene coating on the aluminum heating blocks of the new enFlow intravenous fluid warmer does not negatively affect its performance compared to the uncoated model.
围手术期低体温是一种常见现象,尤其在老年和儿科人群中更为常见。低体温与围手术期并发症的风险增加有关。预防低体温的一种方法是对手术中输注的液体进行加热。enFlow™静脉输液加温器最近重新推出,其加热块涂有聚对二甲苯涂层。在本文中,我们评估了聚对二甲苯涂层对新 enFlow 液体加热能力的影响。
使用 6 个涂覆和 6 个未涂覆的 enFlow 盒。将 10%丙二醇和 90%蒸馏水的溶液以 2、10、50、150 和 200 ml/min 的流速注入每个加热盒中。输注液温度设置为 4°C、20°C 和 37°C。记录每个水平的输出温度。数据分析来自每个流速(三种温度下的六个盒)的 18 次运行。
无论输注液的温度如何,涂覆有聚对二甲苯的液体加温盒在流速为 2、10 和 50 ml/min 时,都能稳定地输出 40°C 的温度。在较高的流速下,对于较冷的液体,盒体无法达到目标温度。在所有温度和流速下,两个盒体的性能相似。
在低流速下,涂覆有聚对二甲苯的 enFlow 盒与原始未涂覆的盒体相当。在较高的流速下,涂覆和未涂覆的盒体都无法达到目标温度。与未涂覆的模型相比,新的 enFlow 静脉输液加温器的铝加热块上的聚对二甲苯涂层不会对其性能产生负面影响。