微孔板中的一步Du试验。

A one-step Du test in microplates.

作者信息

Epley K M, Severns M L, Kline L M

机构信息

American Red Cross, Biomedical Research and Development Laboratories, Rockville, Maryland.

出版信息

Transfusion. 1988 May-Jun;28(3):233-6. doi: 10.1046/j.1537-2995.1988.28388219150.x.

Abstract

A one-step Du test, developed for use in automated microplate systems, uses anti-D with 0.6 percent dextran to potentiate the reaction. Because the washing and reagent-adding steps of the antiglobulin test are not required, the Du test can be performed in the same microplate as the ABO/Rh test. A set of reactions prepared with this technique was visually interpreted and also classified by an automated microplate ABO/Rh system. Visual interpretation of reactions resulted in a sensitivity and specificity close to those of the antiglobulin test, although the sensitivity of the test was reagent-dependent. When the automated microplate blood grouping system was used, the test was not as sensitive or as specific as the antiglobulin test, although it may be sufficient for many applications.