Clin Lab Haematol. 1990;12(4):437-60.
This document is an introduction to microplate serology prepared at a time of very active research and development. Antibody detection techniques are expected to improve during the lifetime of this document. Approximately 25% of hospital blood banks are ABO and D grouping by microplate techniques. Antibody screening by microplate techniques is only being carried out by a small percentage of hospital blood banks but there is likely to be a change to antiglobulin tests by microplate procedures as soon as the methods have been thoroughly tested and approved. There are one or two weak links in these procedures which could give rise to serious errors: (1) Antiglobulin reagents standardized for spin-tube tests may be subject to prozones in liquid phase microplate tests. Each new batch of antiglobulin reagent must be standardized by the microplate procedure in use to show that the reagent is at its optimum anti-IgG dilution for use. However dilutions greater than 1:2 may seriously compromise the anti-complement activity of the reagent. (2) Automated cell washers for microplates are currently under development and should be evaluated by replicate tests with weak anti-D sensitized cells. (3) The combination of diluted anti-IgG and poor washing procedures may lead to neutralization of anti-IgG and cause false negative errors. Future development for improved antiglobulin tests in microplates by a solid phase system is now well advanced and workers proposing to change to antibody screening by microplates are advised to bear this in mind. It is hoped that microplate methodology will move towards standardization based on national guidelines using standard antibody reagents for the validation of new microplate procedures.
本文档是在研发非常活跃的时期编写的关于微孔板血清学的介绍。在本文档的有效期内,抗体检测技术有望得到改进。大约25%的医院血库采用微孔板技术进行ABO和D血型分组。只有一小部分医院血库采用微孔板技术进行抗体筛查,但一旦这些方法经过充分测试和批准,很可能会改用微孔板程序进行抗球蛋白试验。这些程序中有一两个薄弱环节可能会导致严重错误:(1)用于试管试验标准化的抗球蛋白试剂在液相微孔板试验中可能会出现前带现象。每一批新的抗球蛋白试剂都必须通过所使用的微孔板程序进行标准化,以表明该试剂处于其最佳抗IgG稀释度以供使用。然而,大于1:2的稀释可能会严重损害试剂的抗补体活性。(2)微孔板自动细胞洗涤器目前正在研发中,应通过使用弱抗-D致敏细胞的重复试验进行评估。(3)稀释的抗IgG与不良洗涤程序的结合可能导致抗IgG中和并造成假阴性错误。目前,通过固相系统改进微孔板抗球蛋白试验的未来发展已经取得了很大进展,建议打算改用微孔板进行抗体筛查的工作人员牢记这一点。希望微孔板方法将朝着基于国家指南的标准化方向发展,使用标准抗体试剂来验证新的微孔板程序。