The purpose of this study is 1) to demonstrate reproducibility of spin echo-echo planar imaging (SE-EPI) magnetic resonance elastography (MRE) to estimate kidney stiffness; and 2) to compare SE-EPI MRE and gradient recalled echo (GRE) MRE-derived stiffness estimations in various anatomical regions of the kidney. Kidney MRE was performed on 33 healthy subjects (8 for SE-EPI MRE reproducibility and 25 for comparison with GRE MRE; age range: 22-66 years) in a 3 T MRI scanner. To demonstrate SE-EPI MRE reproducibility, subjects were scanned for the first scan and then asked to leave the scan room and repositioned again for the second (repeat) scan. Similar set-up was used for GRE MRE as well. The displacement data was then processed to obtain overall stiffness estimates of the kidney. Concordance correlation analyses were performed to determine SE-EPI MRE reproducibility and agreement between GRE MRE and SE-EPI MRE derived stiffness. A high concordance correlation (ρ = 0.95; p-value<0.0001) was obtained for SE-EPI MRE reproducibility. Good concordance correlation was observed (ρ = 0.84; p < 0.0001 for both kidneys, ρ = 0.91; p < 0.0001 for right kidney and ρ = 0.78; p < 0.0001 for left kidney) between GRE MRE and SE-EPI MRE derived stiffness measurements. Paired t-test results showed that stiffness value of medulla was significantly (p < 0.0001) greater than cortex using SE-EPI MRE as well as GRE MRE. SE-EPI MRE was reproducible and good agreement was observed in MRE-derived stiffness measurements obtained using SE-EPI and GRE sequences. Therefore, SE-EPI can be used for kidney MRE applications.
1)验证自旋回波-回波平面成像(SE-EPI)磁共振弹性成像(MRE)在评估肾脏硬度方面的可重复性;2)比较 SE-EPI MRE 和梯度回波(GRE)MRE 在肾脏不同解剖区域的硬度估计值。在 3T MRI 扫描仪上对 33 名健康受试者(8 名用于 SE-EPI MRE 重复性,25 名用于与 GRE MRE 比较;年龄范围:22-66 岁)进行肾脏 MRE。为了证明 SE-EPI MRE 的可重复性,受检者首先进行扫描,然后离开扫描室重新定位进行第二次(重复)扫描。GRE MRE 也采用了类似的设置。然后处理位移数据以获得肾脏整体硬度估计值。采用一致性相关分析来确定 SE-EPI MRE 的可重复性以及 GRE MRE 和 SE-EPI MRE 衍生的硬度之间的一致性。SE-EPI MRE 重复性获得了高度一致的相关性(ρ=0.95;p 值<0.0001)。GRE MRE 和 SE-EPI MRE 衍生的硬度测量值之间观察到良好的一致性相关性(ρ=0.84;p<0.0001 双侧肾脏,ρ=0.91;p<0.0001 右侧肾脏,ρ=0.78;p<0.0001 左侧肾脏)。SE-EPI MRE 检测到肾髓质的硬度值显著(p<0.0001)高于皮质。SE-EPI MRE 和 GRE MRE 均可重复性良好,并且在 SE-EPI 和 GRE 序列获得的 MRE 衍生的硬度测量值中观察到良好的一致性。因此,SE-EPI 可用于肾脏 MRE 应用。