肾脏磁共振弹性成像:SE-EPI MRE 的可重复性及其与 GRE MRE 的比较。

Magnetic Resonance Elastography of kidneys: SE-EPI MRE reproducibility and its comparison to GRE MRE.

机构信息

Department of Biomedical Engineering, The Ohio State University, Columbus, OH.

Department of Radiology, The Ohio State University Wexner Medical Center, Columbus, OH.

出版信息

NMR Biomed. 2019 Nov;32(11):e4141. doi: 10.1002/nbm.4141. Epub 2019 Jul 22.

Abstract

The purpose of this study is 1) to demonstrate reproducibility of spin echo-echo planar imaging (SE-EPI) magnetic resonance elastography (MRE) to estimate kidney stiffness; and 2) to compare SE-EPI MRE and gradient recalled echo (GRE) MRE-derived stiffness estimations in various anatomical regions of the kidney. Kidney MRE was performed on 33 healthy subjects (8 for SE-EPI MRE reproducibility and 25 for comparison with GRE MRE; age range: 22-66 years) in a 3 T MRI scanner. To demonstrate SE-EPI MRE reproducibility, subjects were scanned for the first scan and then asked to leave the scan room and repositioned again for the second (repeat) scan. Similar set-up was used for GRE MRE as well. The displacement data was then processed to obtain overall stiffness estimates of the kidney. Concordance correlation analyses were performed to determine SE-EPI MRE reproducibility and agreement between GRE MRE and SE-EPI MRE derived stiffness. A high concordance correlation (ρ  = 0.95; p-value<0.0001) was obtained for SE-EPI MRE reproducibility. Good concordance correlation was observed (ρ  = 0.84; p < 0.0001 for both kidneys, ρ  = 0.91; p < 0.0001 for right kidney and ρ  = 0.78; p < 0.0001 for left kidney) between GRE MRE and SE-EPI MRE derived stiffness measurements. Paired t-test results showed that stiffness value of medulla was significantly (p < 0.0001) greater than cortex using SE-EPI MRE as well as GRE MRE. SE-EPI MRE was reproducible and good agreement was observed in MRE-derived stiffness measurements obtained using SE-EPI and GRE sequences. Therefore, SE-EPI can be used for kidney MRE applications.

摘要

本研究的目的是

1)验证自旋回波-回波平面成像(SE-EPI)磁共振弹性成像(MRE)在评估肾脏硬度方面的可重复性;2)比较 SE-EPI MRE 和梯度回波(GRE)MRE 在肾脏不同解剖区域的硬度估计值。在 3T MRI 扫描仪上对 33 名健康受试者(8 名用于 SE-EPI MRE 重复性,25 名用于与 GRE MRE 比较;年龄范围:22-66 岁)进行肾脏 MRE。为了证明 SE-EPI MRE 的可重复性,受检者首先进行扫描,然后离开扫描室重新定位进行第二次(重复)扫描。GRE MRE 也采用了类似的设置。然后处理位移数据以获得肾脏整体硬度估计值。采用一致性相关分析来确定 SE-EPI MRE 的可重复性以及 GRE MRE 和 SE-EPI MRE 衍生的硬度之间的一致性。SE-EPI MRE 重复性获得了高度一致的相关性(ρ=0.95;p 值<0.0001)。GRE MRE 和 SE-EPI MRE 衍生的硬度测量值之间观察到良好的一致性相关性(ρ=0.84;p<0.0001 双侧肾脏,ρ=0.91;p<0.0001 右侧肾脏,ρ=0.78;p<0.0001 左侧肾脏)。SE-EPI MRE 检测到肾髓质的硬度值显著(p<0.0001)高于皮质。SE-EPI MRE 和 GRE MRE 均可重复性良好,并且在 SE-EPI 和 GRE 序列获得的 MRE 衍生的硬度测量值中观察到良好的一致性。因此,SE-EPI 可用于肾脏 MRE 应用。

相似文献

引用本文的文献

本文引用的文献

2
Fibrosis imaging: Current concepts and future directions.纤维化影像学:现状与未来方向。
Adv Drug Deliv Rev. 2017 Nov 1;121:9-26. doi: 10.1016/j.addr.2017.10.013. Epub 2017 Nov 20.
4
Magnetic Resonance Imaging of the Fibrotic Kidney.纤维化肾脏的磁共振成像
J Am Soc Nephrol. 2017 Sep;28(9):2564-2570. doi: 10.1681/ASN.2016101089. Epub 2017 Aug 7.
5
Diagnosis and assessment of renal fibrosis: the state of the art.肾纤维化的诊断与评估:当前技术水平
Swiss Med Wkly. 2017 May 11;147:w14442. doi: 10.4414/smw.2017.14442. eCollection 2017.
10
Treatment of Renal Fibrosis-Turning Challenges into Opportunities.肾纤维化的治疗——将挑战转化为机遇
Adv Chronic Kidney Dis. 2017 Mar;24(2):117-129. doi: 10.1053/j.ackd.2016.11.002.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验