Suppr超能文献

Bactec NR - 660系统与Signal系统的比较。

Comparison of Bactec NR-660 and Signal systems.

作者信息

Rimmer K, Cabot M

机构信息

Department of Medical Microbiology, Royal Liverpool Hospital.

出版信息

J Clin Pathol. 1988 Jun;41(6):676-8. doi: 10.1136/jcp.41.6.676.

Abstract

Bactec NR-660, a computerised blood culture system using infrared analysis of microbial generated carbon dioxide, was compared with the Signal system, which detects gaseous pressure (due to bacterial metabolism) by a manometer. Four trials were undertaken: an in vitro evaluation of 99 bacteria in simulated blood cultures, and three prospective comparisons of a total of 2588 paired patient samples. Combined results for bacteria in simulated blood cultures showed a highly significant difference (p less than 0.001) between Bactec NR-660 aerobic medium (6A) and any other phial under test. Detection rates for most bacteria by Signal were on average three times slower than the first Bactec phial (mean delay 58.3 hours). Overall, the systems were not comparable.

摘要

Bactec NR - 660是一种利用微生物产生的二氧化碳进行红外分析的电脑化血培养系统,它与通过压力计检测(因细菌代谢产生的)气压的信号系统进行了比较。进行了四项试验:对模拟血培养中的99种细菌进行体外评估,以及对总共2588对患者样本进行三项前瞻性比较。模拟血培养中细菌的综合结果显示,Bactec NR - 660需氧培养基(6A)与任何其他受试培养瓶之间存在高度显著差异(p小于0.001)。信号系统对大多数细菌的检测速度平均比第一个Bactec培养瓶慢三倍(平均延迟58.3小时)。总体而言,这两个系统无可比性。

相似文献

1
9

引用本文的文献

5
Detection of bloodstream infections in children.儿童血流感染的检测
Eur J Clin Microbiol Infect Dis. 1989 Sep;8(9):815-24. doi: 10.1007/BF02185854.