Suppr超能文献

一种新型自动红外血液培养系统与BACTEC 460系统的临床比较。

Clinical comparison of a new automated infrared blood culture system with the BACTEC 460 system.

作者信息

Jungkind D, Millan J, Allen S, Dyke J, Hill E

出版信息

J Clin Microbiol. 1986 Feb;23(2):262-6. doi: 10.1128/jcm.23.2.262-266.1986.

Abstract

A new blood culture instrument, the BACTEC (Johnston Laboratories, Inc., Towson, Md.) NR-660, which utilizes infrared detection of carbon dioxide from microbial metabolism, was compared with the radiometric BACTEC 460 system. There were 1,554 isolates from 18,785 paired aerobic blood cultures. Of these isolates, 1,303 were isolated from the radiometric 6B medium, and 1,259 were isolated from the NR6A medium (P = 0.06). Analysis of the data indicated no significant differences in recovery when any individual species was considered. When organisms were considered as groups, there were no significant detection differences for gram-negative bacilli, yeasts, or anaerobes. For gram-positive cocci in aerobic medium, the BACTEC 460 detected 84.3% of the total isolates, and the BACTEC NR-660 detected 79.7% (P = 0.04). There were 891 isolates from 13,983 paired anaerobic blood cultures. Of these isolates, 725 were recovered from the radiometric 7D BACTEC medium, and 723 were recovered from the NR7A BACTEC medium (P greater than 0.9). In the anaerobic media there was no significant difference in detection of any organism group, including the gram-positive cocci. When the results of the aerobic and anaerobic media were combined, there was equivalence between the two systems for the detection of gram-positive cocci (P greater than 0.2) and other organism groups. When the ability to detect septic episodes was compared, there was no significant difference for any organism group (P = 0.12). For aerobic media, the mean times for detection were 30.5 and 29.5 h for the BACTEC 460 and NR-660, respectively. For anaerobic media, the mean times for detection were 39.8 and 41.6 h for the BACTEC 460 NR-660, respectively. Compared with the BACTEC 460, the BACTEC NR-660 system had a greater ease of operation, faster test cycle, computerized data base, and equally rapid detection of positive cultures.

摘要

一种新型血培养仪器,即BACTEC(约翰斯顿实验室公司,马里兰州陶森)NR - 660,它利用红外线检测微生物代谢产生的二氧化碳,并与放射性测量的BACTEC 460系统进行了比较。在18785对需氧血培养中分离出1554株菌株。在这些分离菌株中,1303株从放射性测量的6B培养基中分离得到,1259株从NR6A培养基中分离得到(P = 0.06)。数据分析表明,考虑任何单个菌种时,回收率没有显著差异。当将微生物作为群体考虑时,革兰氏阴性杆菌、酵母菌或厌氧菌的检测没有显著差异。对于需氧培养基中的革兰氏阳性球菌,BACTEC 460检测到总分离菌株的84.3%,BACTEC NR - 660检测到79.7%(P = 0.04)。在13983对厌氧血培养中分离出891株菌株。在这些分离菌株中,725株从放射性测量的BACTEC 7D培养基中回收得到,723株从NR7A BACTEC培养基中回收得到(P大于0.9)。在厌氧培养基中,任何微生物群体(包括革兰氏阳性球菌)的检测没有显著差异。当将需氧和厌氧培养基的结果合并时,两种系统在检测革兰氏阳性球菌(P大于0.2)和其他微生物群体方面相当。当比较检测败血症发作的能力时,任何微生物群体都没有显著差异(P = 0.12)。对于需氧培养基,BACTEC 460和NR - 660的平均检测时间分别为30.5小时和29.5小时。对于厌氧培养基,BACTEC 460和NR - 660的平均检测时间分别为39.8小时和41.6小时。与BACTEC 460相比,BACTEC NR - 660系统操作更简便,测试周期更快,有计算机化数据库,且阳性培养物检测同样迅速。

相似文献

7
Routine evaluation of the nonradiometric BACTEC NR 660 system.非放射性BACTEC NR 660系统的常规评估
J Clin Microbiol. 1986 Jul;24(1):26-9. doi: 10.1128/jcm.24.1.26-29.1986.

引用本文的文献

2
Gram-negative bacteremia in non-neutropenic patients: a 3-year review.
Infection. 1998 May-Jun;26(3):155-9. doi: 10.1007/BF02771841.
3
Update on detection of bacteremia and fungemia.菌血症和真菌血症检测的最新进展。
Clin Microbiol Rev. 1997 Jul;10(3):444-65. doi: 10.1128/CMR.10.3.444.
4
Detection of bacteremia by Difco ESP blood culture system.使用Difco ESP血培养系统检测菌血症
J Clin Microbiol. 1994 Mar;32(3):811-8. doi: 10.1128/jcm.32.3.811-818.1994.
5
Current controversies in the detection of septicemia.败血症检测中的当前争议
Eur J Clin Microbiol. 1986 Oct;5(5):487-91. doi: 10.1007/BF02017688.
6
Routine evaluation of BACTEC NR-16A and NR-17A media.BACTEC NR - 16A和NR - 17A培养基的常规评估。
J Clin Microbiol. 1988 Sep;26(9):1619-22. doi: 10.1128/jcm.26.9.1619-1622.1988.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验