BACTEC 9050和BACTEC 9240血培养系统的多中心比较。

Multicenter comparison of BACTEC 9050 and BACTEC 9240 blood culture systems.

作者信息

Murray P R, Hollick G E, Jerris R C, Wilson M L

机构信息

Division of Laboratory Medicine, Washington University, St. Louis, Missouri 63110, USA.

出版信息

J Clin Microbiol. 1998 Jun;36(6):1601-3. doi: 10.1128/JCM.36.6.1601-1603.1998.

Abstract

The overall recovery of organisms and time to detection with the BACTEC 9050 and BACTEC 9240 systems were compared in a multicenter evaluation. In the first phase of the study, a total of 4,383 compliant aerobic (Plus Aerobic/F) blood culture sets were processed. There was no significant difference in the recovery of individual groups of organisms with the two systems, with the exception of Streptococcus pneumoniae which was isolated more frequently with BACTEC 9050. False-positive signals occurred more often with BACTEC 9240 (58 cultures) than with BACTEC 9050 (43 cultures), but false-negative cultures were uncommon with both systems (3 cultures for each system). Time to detection of positive cultures of clinically significant organisms was essentially the same with both instruments. In the second phase of the study, 2,431 compliant anaerobic (Plus Anaerobic/F) blood culture sets were processed. There was no significant difference in the recovery of organisms with BACTEC 9050 compared with BACTEC 9240. Significantly (P < 0.03) more false-positive signals occurred with BACTEC 9240 (15 cultures) than with BACTEC 9050 (4 cultures). Likewise, more false-negative cultures occurred with BACTEC 9240 (11 cultures) than with BACTEC 9050 (8 cultures). Time to detection of positive cultures of clinically significant organisms was essentially the same with both systems with the exception of anaerobes (N = 10), which were recovered earlier (P < 0.01) with BACTEC 9240 (35.0 h) than with BACTEC 9050 (61.4 h).

摘要

在一项多中心评估中,对使用BACTEC 9050和BACTEC 9240系统时微生物的总体回收率及检测时间进行了比较。在研究的第一阶段,共处理了4383套符合要求的需氧(加需氧/F)血培养瓶。除肺炎链球菌在BACTEC 9050系统中分离频率更高外,两种系统对各微生物组的回收率无显著差异。BACTEC 9240系统出现假阳性信号(58瓶培养)的频率高于BACTEC 9050系统(43瓶培养),但两种系统的假阴性培养均不常见(各系统均为3瓶培养)。两种仪器对临床意义重大的微生物阳性培养的检测时间基本相同。在研究的第二阶段,处理了2431套符合要求的厌氧(加厌氧/F)血培养瓶。与BACTEC 9240相比,BACTEC 9050对微生物的回收率无显著差异。BACTEC 9240系统出现假阳性信号(15瓶培养)的频率显著高于BACTEC 9050系统(4瓶培养)(P<0.03)。同样,BACTEC 9240系统出现假阴性培养(11瓶培养)的频率高于BACTEC 9050系统(8瓶培养)。两种系统对临床意义重大的微生物阳性培养的检测时间基本相同,但厌氧菌除外(N=10),BACTEC 9240系统(35.0小时)比BACTEC 9050系统(61.4小时)更早检出厌氧菌(P<0.01)。

相似文献

8
Detection of bacteremia by Difco ESP blood culture system.使用Difco ESP血培养系统检测菌血症
J Clin Microbiol. 1994 Mar;32(3):811-8. doi: 10.1128/jcm.32.3.811-818.1994.

本文引用的文献

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验