Implant Dent. 2019 Aug;28(4):372-377. doi: 10.1097/ID.0000000000000908.
Steam heat sterilization could be one of the factors that affects the dimensional accuracy of surgical guides, leading to an error during guided implant surgery.
The purpose of this study was to investigate the effect of steam heat sterilization on the dimensional changes of surgical guides.
A total of 27 surgical guides, which were made by either Formlabs printer (Formlabs Inc., Somerville, MA) or Simplant (DENTSPLY Implants NV, Hasselt, Belgium), were scanned using an intraoral scanner before and after sterilization. The dimensional changes at the center of the implant sleeves were analyzed using the computer-aided design interactive software for 65 implant sites before and after steam heat sterilization at 121°C for 20 minutes.
There was no significant difference between the mean x, y, and z axes of the center of the sleeves when measured before and after sterilization (P values were 0.37, 0.24, and 0.29, respectively). Nonparametric analysis showed no significant difference between the mean deviations of either surgical guide (P = 0.908).
Steam heat sterilization has a nonsignificant effect on the dimensional changes of the tested surgical guides.
蒸汽热消毒可能是影响手术导板尺寸精度的因素之一,从而导致引导种植手术中的误差。
本研究旨在探讨蒸汽热消毒对手术导板尺寸变化的影响。
共扫描了 27 个手术导板,分别由 Formlabs 打印机(Formlabs Inc.,马萨诸塞州萨默维尔)或 Simplant(登士柏种植体有限公司,比利时哈塞尔特)制作,在消毒前后使用口腔内扫描仪进行扫描。在 121°C 下蒸汽热消毒 20 分钟后,使用计算机辅助设计交互软件分析 65 个种植体部位的种植套管中心的尺寸变化。
消毒前后套管中心的 x、y 和 z 轴的平均轴无显著差异(P 值分别为 0.37、0.24 和 0.29)。非参数分析显示,两种手术导板的平均偏差无显著差异(P = 0.908)。
蒸汽热消毒对测试手术导板的尺寸变化无显著影响。