灭菌对3D打印手术导板精度和贴合度的影响。

The Effect of Sterilization on the Accuracy and Fit of 3D-Printed Surgical Guides.

作者信息

Yazigi Christine, Chaar M Sad, Busch Reinhard, Kern Matthias

机构信息

Department of Prosthodontics, Propaedeutics and Dental Materials, School of Dentistry, Christian-Albrechts University, 24105 Kiel, Germany.

出版信息

Materials (Basel). 2023 Jul 28;16(15):5305. doi: 10.3390/ma16155305.

Abstract

This study was conducted to evaluate the accuracy of 3D-printed surgical guides before and after sterilization in a steam sterilizer. A test-model incorporating three implant replicas was customized. A total of forty guides were printed from five printable resins. A group made from a self-curing composite served as control group. The guides were checked for fit. Vertical discrepancies between the model and guides were measured at standardized points at a load of 500 g (P1). The guides were connected to implant replicas and scanned, and their angles were digitally measured. The specimens were sterilized in a steam sterilizer at 121 °C for 20 min at 2 bar pressure. Vertical discrepancies (P2) and angulations were remeasured. Additionally, the specimens were repositioned with an increased load, and measurements were repeated (P3). All specimens were repositionable after sterilization. The smallest variation in discrepancy at a 500 g load was 428 µm, whereas the greatest was 1487 µm. Under an increased force, the smallest change was 94 µm, while the greatest was 260 µm. The level of significance α = 0.05 (95% confidence interval) was set for all tests. The variation in the measured angles was not statistically significant (Kruskal-Wallis's test, > 0.05). The accuracy was affected by the material and sterilization, but it was clinically acceptable when an increased load was applied during repositioning.

摘要

本研究旨在评估3D打印手术导板在蒸汽灭菌器灭菌前后的准确性。定制了一个包含三个种植体复制体的测试模型。用五种可打印树脂共打印了40个导板。由自固化复合材料制成的一组作为对照组。检查导板的适配情况。在500 g负荷(P1)下,在标准化点测量模型与导板之间的垂直差异。将导板与种植体复制体连接并扫描,数字测量其角度。标本在蒸汽灭菌器中于121℃、2巴压力下灭菌20分钟。重新测量垂直差异(P2)和角度。此外,增加负荷重新放置标本并重复测量(P3)。所有标本灭菌后均可重新定位。500 g负荷下差异的最小变化为428 µm,最大变化为1487 µm。在增加力的情况下,最小变化为94 µm,最大变化为260 µm。所有测试设定显著性水平α = 0.05(95%置信区间)。测量角度的变化无统计学意义(Kruskal-Wallis检验,>0.05)。准确性受材料和灭菌的影响,但在重新定位时增加负荷的情况下,临床上是可接受的。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/89c7/10419648/1c1960bf5ad5/materials-16-05305-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验