: The purpose of this study is to determine whether the Sbisa bar (Sbisa Ophthalmic Instruments) (referred to in this article as the BF(Sbisa) bar) and Bagolini Filter bar (C.O.I. Vision) (referred to in this article as the BF(new) bar) are repeatable and comparable, for measuring density of suppression or strength of abnormal binocular single vision (BSV), in a pediatric population. : Using a spot photometer, percentage light transmittance of filters on two BF(Sbisa) bars and two BF(new) bars was measured. Thirty-three participants aged 5-12 years with unilateral strabismus were recruited. Density of suppression or strength of abnormal BSV was measured with two of each bar, in addition to a repeated measure with one of each bar. : Light transmittance did not reduce consistently with increasing filter number (the number assigned to each filter by the manufacturer). Using data as filter numbers, a BF(Sbisa) bar was not comparable to a BF(new) bar ( = -3.936, < 0.001, = -0.28); different BF(Sbisa) bars were not comparable ( = -3.103, = 0.005, = -0.22) but different BF(new) bars were comparable ( = -1.165, = 1.285, = -0.08). Using light transmittance values, a BF(Sbisa) bar was comparable to a BF(new) bar ( = -0.385, = 3.535, = -0.03); different BF(Sbisa) bars were comparable ( = -1.476, = 0.720, = -0.10) but different BF(new) bars were not comparable ( = -3.354, < 0.001, = -0.24). The BF(Sbisa) and BF(new) bars were not repeatable using filter numbers ( = -3.648, < 0.001, = -0.26 and = -3.099, = 0.005, = -0.22, respectively) or light transmittance values ( = -3.743, < 0.001, = -0.27 and = -2.727, = 0.025, = -0.19, respectively). An order effect existed using filter numbers ( = -4.473, < 0.001, = -0.32) and light transmittance values ( = -4.525, < 0.001, = -0.32). : There is a need for the production of bars with consistently reducing and comparable light transmittance of filters if measurement of density of suppression or strength of abnormal BSV with filter bars is to continue, because the current bars show inconsistencies, particularly at lower filter numbers. In the meantime, the bars are limited by their lack of comparability or repeatability. Using only BF(new) bars is preferable to using BF(Sbisa) bars or a mixture of the two types of bar because two different BF(new) bars may be comparable using filter number readings, but filter number readings from different BF(Sbisa) bars or from a BF(Sbisa) and BF(new) bar are not comparable. As the BF(Sbisa) and BF(new) bars are currently labeled, a second measure of density of suppression or strength of abnormal BSV is not recommended within 2 min of an initial measurement, because the measurement procedure seems to cause a reduction in suppression or abnormal BSV that does not recover with repeated measurements 2 min apart.
本研究的目的是确定Sbisa条(Sbisa眼科仪器)(本文中称为BF(Sbisa)条)和Bagolini滤光条(C.O.I.视觉)(本文中称为BF(new)条)在测量儿童群体的抑制密度或异常双眼单视(BSV)强度时是否具有可重复性和可比性。:使用点光度计测量了两根BF(Sbisa)条和两根BF(new)条上滤光片的透光率百分比。招募了33名年龄在5至12岁的单侧斜视参与者。除了对每种条各重复测量一次外,还用每种条各两根测量了抑制密度或异常BSV的强度。:透光率并未随着滤光片数量(制造商为每个滤光片指定的编号)的增加而持续降低。以滤光片编号作为数据时,BF(Sbisa)条与BF(new)条不可比(t = -3.936,p < 0.001,r = -0.28);不同的BF(Sbisa)条不可比(t = -3.103,p = 0.005,r = -0.22),但不同的BF(new)条具有可比性(t = -1.165,p = 1.285,r = -0.08)。以透光率值作为数据时,BF(Sbisa)条与BF(new)条具有可比性(t = -0.385,p = 3.535,r = -0.03);不同的BF(Sbisa)条具有可比性(t = -1.476,p = 0.720,r = -0.10),但不同的BF(new)条不可比(t = -3.354,p < 0.001,r = -0.24)。使用滤光片编号或透光率值时,BF(Sbisa)条和BF(new)条均无重复性(t分别为 -3.648,p < 0.001,r = -0.26和t = -3.099,p = 0.005,r = -0.22)。使用滤光片编号(t = -4.473,p < 0.001,r = -0.32)和透光率值(t = -4.525,p < 0.001,r = -0.32)时存在顺序效应。:如果要继续使用滤光条测量抑制密度或异常BSV强度,就需要生产滤光片透光率持续降低且具有可比性的条,因为目前的条存在不一致性,尤其是在滤光片编号较低时。同时,这些条受到缺乏可比性或可重复性的限制。仅使用BF(new)条比使用BF(Sbisa)条或两种条混合使用更可取,因为使用滤光片编号读数时,两根不同的BF(new)条可能具有可比性,但来自不同BF(Sbisa)条或BF(Sbisa)条与BF(new)条的滤光片编号读数不可比。按照目前BF(Sbisa)条和BF(new)条的标记方式,不建议在首次测量后2分钟内对抑制密度或异常BSV强度进行第二次测量,因为测量过程似乎会导致抑制或异常BSV降低,且间隔2分钟重复测量也无法恢复。