Mastropasqua Leonardo, Toto Lisa, D'Ugo Erminia, Lanzini Manuela, Mattei Peter A, Falconio Gennaro, Doronzo Emanuele, Diomede Francesca, Trubiani Oriana
Department of Medicine and Science of Ageing, Ophthalmology Clinic, University G. D'Annunzio Chieti-Pescara, Via dei Vestini, 66100, Chieti, Italy.
Laboratory of Stem Cells and Regenerative Medicine, Department of Medical, Oral and Biotechnological Sciences,, University G. D'Annunzio Chieti-Pescara, Chieti, Italy.
Int Ophthalmol. 2020 Jan;40(1):125-134. doi: 10.1007/s10792-019-01154-0. Epub 2019 Aug 26.
To compare the corneal tissue trauma after the use of an automated preloaded injector and a manual injector and assess scanning electron microscope (SEM) and atomic force microscope (AFM) features of both injector cartridges.
Ophthalmology Clinic and Laboratory of Stem Cells and Regenerative Medicine University "G. d'Annunzio" of Chieti-Pescara, Chieti, Italy; DESIGN: Prospective randomized clinical study METHODS: Forty eyes of 40 patients for phacoemulsification were divided into two groups: implantation of intraocular lens was performed with AutonoMe automated delivery system (AutonoMe group: 20 eyes) and Monarch III injector system (Monarch group: 20 eyes). In vivo confocal microscopy (IVCM) and anterior segment optical coherence tomography (AS-OCT) were performed before surgery, at 1 h, 1 day and 1 month post-operatively. In addition, SEM and AFM were performed on cartridges of both injector systems after injection of the IOL.
A greater increase in central corneal thickness and corneal thickness at the incision site were observed in Monarch group versus AutonoMe group 1 h and 1 day post-operatively (p < 0.05). Endothelial cell count loss was significantly higher in Monarch group compared with AutonoMe group (p < 0.05) at 1 and 24 h. AS-OCT showed less endothelial misalignment at 30 days (p < 0.05), and IVCM showed less tunnel inflammation at all time points (p < 0.05) in AutonoMe group compared with Monarch group; roughness analysis at AFM of the AutonoMe cartridge was significantly lower compared to Monarch D cartridge (p < 0.05).
The AutonoMe injector provided less corneal tissue trauma compared with Monarch III injector. The AutonoMe cartridge showed lower roughness at AFM compared to the Monarch D cartridge.
比较使用自动预装注射器和手动注射器后角膜组织的损伤情况,并评估两种注射器药筒的扫描电子显微镜(SEM)和原子力显微镜(AFM)特征。
意大利基耶蒂-佩斯卡拉“G. d'Annunzio”大学干细胞与再生医学眼科诊所及实验室
前瞻性随机临床研究
40例接受白内障超声乳化手术患者的40只眼被分为两组:分别使用AutonoMe自动输送系统(AutonoMe组:20只眼)和Monarch III注射器系统(Monarch组:20只眼)植入人工晶状体。术前、术后1小时、1天和1个月进行活体共聚焦显微镜检查(IVCM)和眼前节光学相干断层扫描(AS-OCT)。此外,在注入人工晶状体后,对两种注射器系统的药筒进行SEM和AFM检查。
术后1小时和1天,Monarch组中央角膜厚度和切口部位角膜厚度的增加幅度大于AutonoMe组(p < 0.05)。术后1小时和24小时,Monarch组内皮细胞计数损失显著高于AutonoMe组(p < 0.05)。AS-OCT显示,术后30天AutonoMe组内皮细胞排列不齐情况少于Monarch组(p < 0.05),IVCM显示,在所有时间点AutonoMe组隧道炎症均少于Monarch组(p < 0.05);AutonoMe药筒的AFM粗糙度分析显著低于Monarch D药筒(p < 0.05)。
与Monarch III注射器相比,AutonoMe注射器造成的角膜组织损伤更小。与Monarch D药筒相比,AutonoMe药筒的AFM粗糙度更低。