Usui Masahiko, Tanaka Takao
From Tokyo Medical University (Usui) and the Eye Institute (Tanaka), Ebisu Clinic Ganka, Tokyo, Japan.
From Tokyo Medical University (Usui) and the Eye Institute (Tanaka), Ebisu Clinic Ganka, Tokyo, Japan.
J Cataract Refract Surg. 2015 Aug;41(8):1745-51. doi: 10.1016/j.jcrs.2015.03.018.
To measure and compare the resistance force for intraocular lens (IOL) insertion using 5 syringe-type injector systems.
Tokyo Medical University and laboratory in Kowa Co., Tokyo, Japan.
Experimental study.
Intraocular lenses were inserted into the lens capsular bag of porcine eyes after phacoemulsification using 5 implantation systems (Groups A, B, C, D, and E). For each system, the resistance force for IOL insertion to the lens capsular bag of porcine eyes was measured using an automated force gauge system. For control, the resistance force for IOL delivery into a plastic dish was measured. Changes in the resistance force and its curve and maximum value were evaluated. The mean total area under the curve (AUC) was compared. Data were statistically analyzed.
For all groups, the mean resistance forces were 17.2, 6.3, 4.2, 20.7, and 2.3 newtons (N), respectively, in porcine eyes, and 14.4, 5.8, 4.5, 12.6, and 2.2 N in controls. The mean sizes of the total AUC were 43 371, 8465, 6771, 30 306, and 2334 pixels in porcine eyes and 40 940, 7080, 6876, 20 710, and 2215 pixels in controls; the correlation coefficients between the resistance forces and the sizes of the total area were 0.576, 0.113, 0.346, 0.726, and 0.933 in porcine eyes and 0.707, 0.557, 0.914, 0.951, and 0.893 in controls.
Resistance force and its curve were clarified in 5 IOL implantation systems. Appropriate IOL and injector selection may be achieved after clarifying resistance force and its waveform during IOL insertion.
Neither author has a financial or proprietary interest in any material or method mentioned.
使用5种注射器式注入系统测量并比较人工晶状体(IOL)植入时的阻力。
东京医科大学以及日本东京兴和公司的实验室。
实验研究。
使用5种植入系统(A组、B组、C组、D组和E组)在超声乳化术后将人工晶状体植入猪眼的晶状体囊袋内。对于每种系统,使用自动测力计系统测量人工晶状体植入猪眼晶状体囊袋时的阻力。作为对照,测量人工晶状体注入塑料培养皿时的阻力。评估阻力及其曲线和最大值的变化。比较曲线下的平均总面积(AUC)。对数据进行统计学分析。
在所有组中,人工晶状体植入猪眼时的平均阻力分别为17.2、6.3、4.2、20.7和2.3牛顿(N),对照组中分别为14.4、5.8、4.5、12.6和2.2 N。猪眼中总AUC的平均大小分别为43371、8465、6771、30306和2334像素,对照组中分别为40940、7080、6876、20710和2215像素;猪眼中阻力与总面积大小之间的相关系数分别为0.576、0.113、0.346、0.726和0.933,对照组中分别为0.707、0.557、0.914、0.951和0.893。
明确了5种人工晶状体植入系统中的阻力及其曲线。在明确人工晶状体植入时的阻力及其波形后,可能实现人工晶状体和注入器的适当选择。
两位作者均对文中提及的任何材料或方法没有财务或专有利益。