用于检查全身计数器和活化分析系统校准的人体模型。

Anthropomorphic models for checking the calibration of whole-body counters and activation analysis systems.

作者信息

Bewley D K

机构信息

MRC Cyclotron Unit, Hammersmith Hospital, London, UK.

出版信息

Phys Med Biol. 1988 Jul;33(7):805-13. doi: 10.1088/0031-9155/33/7/004.

Abstract

Three models of the human body have been made for checking the calibration of whole-body counters and activation analysis systems. Their weights are 41, 67 and 110 kg. Each contains simplified forms of lungs, bones, thyroid, kidneys and liver and has a chemical composition similar to that of the ICRP standard man. In addition traces of toxic elements such as Cd and Hg have been included in various organs. The composition of the manikins has been measured at four centres where activation analysis in vivo has been carried out. The potassium contents have been measured at three other centres. Differences between measured and actual contents have often been much larger than quoted uncertainties of the measurements, 20-50% for Na, Cl and P. For K, differences did not exceed 17% but are still often greater than the expected uncertainty. The greatest errors usually occurred with the smallest manikin.

摘要

已制作了三种人体模型,用于检查全身计数器和活化分析系统的校准。它们的重量分别为41、67和110千克。每个模型都包含简化形式的肺、骨骼、甲状腺、肾脏和肝脏,其化学成分与国际辐射防护委员会(ICRP)标准人体相似。此外,各个器官中还含有微量的镉和汞等有毒元素。这些人体模型的成分已在四个进行体内活化分析的中心进行了测量。钾含量在另外三个中心进行了测量。测量值与实际值之间的差异通常远大于所引用的测量不确定度,钠、氯和磷的差异为20 - 50%。对于钾,差异不超过17%,但仍常常大于预期的不确定度。最大的误差通常出现在最小的人体模型上。