在不同钻速下,评估密质人工骨块中窄直径种植体稳定性的共振频率分析装置的体外比较。

In vitro comparison of resonance frequency analysis devices to evaluate implant stability of narrow diameter implants at varying drilling speeds in dense artificial bone blocks.

机构信息

Department of Periodontology, Laboratory for Periodontal-, Implant-, Phototherapy (LA-PIP), School of Dental Medicine, Stony Brook University, Stony Brook, New York.

Department of Oral Surgery and Implant Dentistry, Dental School, Johann Wolfgang Goethe University, Frankfurt, Germany.

出版信息

Clin Implant Dent Relat Res. 2019 Oct;21(5):1023-1027. doi: 10.1111/cid.12842. Epub 2019 Aug 29.

Abstract

BACKGROUND

There are no studies that have assessed the implant stability quotient (ISQ) values of narrow diameter implants placed in artificial dense bone blocks at varying drilling speeds (DSs).

PURPOSE

The aim of the present in vitro experiment was to compare the performance of OSSTELL and Penguin devices to evaluate implant stability at DSs of 800 and 2000 rpm.

MATERIALS AND METHODS

A total of 360 osteotomies were created in dense artificial bone blocks at DSs of 800 and 2000 rpm. Dental implants from three manufacturers (group-1: NobelActive implants, Nobel Biocare, Yorba Linda, California; group-2: Zimmer, Eztetic-Zimmer implants, Zimmer Biomet Dental, Palm Beach Gardens, Florida; and group-3: Astra Tech implant system, Dentsply Sirona, York, Pennsylvania) were randomly placed in these osteotomies using an insertion torque of 15 Ncm (60 implants/group). Implant stability in all bone blocks immediately following implant placement was evaluated using the OSSTELL and Penguin devices. ISQ values were presented as means ± SD. Statistical significance was set at P < .05.

RESULTS

There was no significant difference in the ISQ values obtained from the OSSTELL and Penguin devices for implants in groups 1, 2, and 3. There was no significant difference when ISQ values obtained from the OSSTELL device were compared with the Penguin device for narrow diameter dental implants placed in dense bone blocks with osteotomies performed at 800 and 2000 rpm. ISQ values showed statistically significant higher values for OSSTELL compared to Penguin device.

CONCLUSION

The OSSTELL and Penguin devices are reliable for the assessment of implant stability in dense artificial bone. Implant design and site-DS does not seem to have a significant impact of implant stability in artificial dense bone blocks.

摘要

背景

目前尚无研究评估在不同钻速(DS)下放置在人工致密骨块中的小直径种植体的种植体稳定性系数(ISQ)值。

目的

本体外实验旨在比较 OSSTELL 和 Penguin 设备的性能,以评估在 800 和 2000rpm 钻速下的种植体稳定性。

材料和方法

在 DS 为 800 和 2000rpm 的致密人工骨块中共制备 360 个骨切。将来自三个制造商的种植体(组 1:NobelActive 种植体,诺贝尔生物,加利福尼亚州约巴林达;组 2:Zimmer,Eztetic-Zimmer 种植体, Zimmer Biomet Dental,佛罗里达州棕榈滩花园;组 3:Astra Tech 种植体系统,登士柏西诺德,宾夕法尼亚州约克)随机放置在这些骨切中,使用 15Ncm 的插入扭矩(每组 60 个种植体)。在种植体放置后立即使用 OSSTELL 和 Penguin 设备评估所有骨块中的种植体稳定性。ISQ 值以平均值 ± SD 表示。统计学意义设为 P<.05。

结果

在组 1、2 和 3 中,从 OSSTELL 和 Penguin 设备获得的 ISQ 值之间没有显著差异。在 800 和 2000rpm 下进行骨切时,将从 OSSTELL 设备获得的 ISQ 值与 Penguin 设备进行比较时,窄直径种植体在致密骨块中的 ISQ 值也没有显著差异。与 Penguin 设备相比,OSSTELL 设备的 ISQ 值显示出统计学上更高的值。

结论

OSSTELL 和 Penguin 设备可用于评估致密人工骨中的种植体稳定性。种植体设计和部位-DS 似乎对人工致密骨块中的种植体稳定性没有显著影响。