Suppr超能文献

虚拟星击分析提供亚毫米级辐射等中心精度的可行性:一项长期的多机构分析。

Feasibility of virtual starshot analysis providing submillimeter radiation isocenter accuracy: A long-term multi-institutional analysis.

机构信息

Oncology Center, Osaka University Hospital, Suita, Osaka, 565-0871, Japan.

Suita Tokushukai Hospital, Suita, Osaka, 565-0814, Japan.

出版信息

J Appl Clin Med Phys. 2019 Oct;20(10):74-83. doi: 10.1002/acm2.12715. Epub 2019 Sep 9.

Abstract

PURPOSE

We developed a technique to calculate the offset between room lasers and the radiation isocenter using a digital Winston-Lutz (WL) test with a starshot technique. We have performed isocenter localization quality assurance (QA) with submillimeter accuracy for a long period. Here we evaluated the feasibility and accuracy of this virtual starshot (VS) analysis for isocenter localization QA.

METHODS

A 6-MV photon beam with a square multileaf collimator field was used to irradiate a WL sphere positioned at the intersection of the room lasers. Images were acquired using an electronic portal imaging device. A four-field WL test was performed, and the path of each beam was calculated from the offset between the beam and sphere. Virtual starshot analysis was used to analyze the radiation isocenter, which calculates the center of the beam paths by using a least-squares method, similar to the starshot analysis. Then, eight coplanar and 12 noncoplanar beams were irradiated to evaluate isocenter localization accuracy.

RESULTS

Several VS analyses, using different WL spheres, were performed at three institutions, and the calculated accuracies were within 0.1 mm at all institutions. Long-term analysis showed that the isocenter localization accuracy was appropriately managed with three-dimensional accuracy within ± 0.5 mm for 90 months after the first laser adjustments. The offset between each beam and the room laser was within 0.6 mm and within 1.0 mm for eight coplanar and 12 noncoplanar beams, respectively, for 90 months. Cone-beam computed tomography images, acquired after verification beams, showed that the offset between the radiation isocenter and the imaging center was within 0.66 mm for 90 months. The isocenter localization accuracy within 1 mm was kept for long period at other four institutions.

CONCLUSIONS

Long-term analysis showed the feasibility of VS analysis for isocenter localization QA, including room laser re-alignment, noncoplanar irradiation verification, and image guidance accuracy.

摘要

目的

我们开发了一种使用带有星点技术的数字 Winston-Lutz(WL)测试来计算房间激光和辐射等中心之间的偏移的技术。我们已经使用亚毫米精度进行了很长时间的等中心定位质量保证(QA)。在这里,我们评估了这种虚拟星点(VS)分析用于等中心定位 QA 的可行性和准确性。

方法

使用带有方形多叶准直器场的 6MV 光子束照射位于房间激光交点处的 WL 球体。使用电子射野影像装置采集图像。进行了四个场的 WL 测试,并且从光束和球体之间的偏移计算出每个光束的路径。虚拟星点分析用于分析辐射等中心,它通过最小二乘法计算光束路径的中心,类似于星点分析。然后,照射了八个共面和十二个非共面光束以评估等中心定位精度。

结果

在三个机构进行了几次 VS 分析,使用了不同的 WL 球体,在所有机构中的计算精度都在 0.1 毫米以内。长期分析表明,在首次激光调整后的 90 个月内,通过三维精度在±0.5 毫米以内,适当管理了等中心定位精度。每个光束与房间激光之间的偏移在 0.6 毫米以内,对于八个共面和十二个非共面光束,分别为 0.6 毫米和 1.0 毫米,在 90 个月内。验证光束后的锥形束计算机断层扫描图像显示,在 90 个月内,辐射等中心和成像中心之间的偏移在 0.66 毫米以内。其他四个机构在 1 毫米以内的等中心定位精度保持了很长时间。

结论

长期分析表明,VS 分析对于等中心定位 QA 包括房间激光重新对准、非共面照射验证和图像引导精度是可行的。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/6747/6806479/f34028cf1f17/ACM2-20-74-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验