Europace. 2019 Jul 1;21(7):1031-1038. doi: 10.1093/europace/euz060.
Atrial fibrillation (AF) is the most common tachyarrhythmia and a significant cause of cardioembolic strokes. Atrial fibrillation is often intermittent and asymptomatic making detection a major clinical challenge. We evaluated a photoplethysmography (PPG) wrist band in individual pulse detection in patients with AF and tested the reliability of two commonly used algorithms for AF detection.
A 5-min PPG was recorded from patients with AF or sinus rhythm (SR) with a wrist band and analysed with two AF detection algorithms; AFEvidence and COSEn. Simultaneously registered electrocardiogram served as the golden standard for rhythm analysis and was interpreted by two cardiologists. The study population consisted of 213 (106 AF, 107 SR) patients. The wrist band PPG achieved individual pulse detection with a sensitivity of 91.7 ± 11.2% and a positive predictive value (PPV) of 97.5 ± 4.6% for AF, with a sensitivity of 99.4 ± 1.5% [7.7% (95% confidence interval, 95% CI 5.5% to 9.9%); P < 0.001] and PPV of 98.1 ± 4.1% [0.6% (95% CI -0.6% to 1.7%); P = 0.350] for SR. The pulse detection sensitivity was lower 86.7 ± 13.9% with recent-onset AF (AF duration <48 h, n = 43, 40.6%) as compared to late AF (≥48 h, n = 63, 59.4%) with 95.1 ± 7.2% [-8.3% (95% CI -12.9% to -3.7%); P = 0.001]. For the detection of AF from the wrist band PPG, the sensitivities were 96.2%/95.3% and specificity 98.1% with two algorithms.
The wrist band PPG enabled accurate algorithm-based detection of AF with two AF detection algorithms and high individual pulse detection. Algorithms allowed accurate detection of AF from the PPG. A PPG wrist band provides an easy solution for AF screening.
心房颤动(AF)是最常见的快速性心律失常,也是心源性脑栓塞的重要原因。AF 常呈间歇性和无症状性,这使得检测成为一项重大的临床挑战。我们评估了一种光电容积脉搏波(PPG)腕带在 AF 患者的个体脉搏检测中的应用,并测试了两种常用于 AF 检测的算法的可靠性。
使用腕带记录了 213 例 AF 或窦性节律(SR)患者的 5 分钟 PPG,并使用两种 AF 检测算法(AF Evidence 和 COSEn)进行分析。同时记录的心电图作为节律分析的金标准,由两名心脏病专家进行解读。研究人群包括 106 例 AF 患者和 107 例 SR 患者。腕带 PPG 实现了个体脉搏检测,其对 AF 的敏感性为 91.7±11.2%,阳性预测值(PPV)为 97.5±4.6%,对 SR 的敏感性为 99.4±1.5%[7.7%(95%置信区间,95%CI 5.5%至 9.9%);P<0.001],PPV 为 98.1±4.1%[0.6%(95%CI -0.6%至 1.7%);P=0.350]。与晚期 AF(≥48 小时,n=63,59.4%)相比,近期发作的 AF(AF 持续时间<48 小时,n=43,40.6%)的脉搏检测敏感性较低,为 86.7±13.9%,分别为 95.1±7.2%[-8.3%(95%CI -12.9%至 -3.7%);P=0.001]。对于从腕带 PPG 检测 AF,两种算法的敏感性分别为 96.2%/95.3%和特异性 98.1%。
腕带 PPG 结合两种 AF 检测算法,能够实现 AF 的准确、基于算法的检测,具有较高的个体脉搏检测能力。算法允许从 PPG 准确检测 AF。PPG 腕带为 AF 筛查提供了一种简单的解决方案。