Orthopaedic Research Centre Amsterdam, Department of Orthopaedics of the Amsterdam University Medical Centre, University of Amsterdam, The Netherlands.
Department of Biomedical Engineering and Physics of the Amsterdam University Medical Centre, University of Amsterdam, The Netherlands.
Hip Int. 2021 Jan;31(1):58-65. doi: 10.1177/1120700019874822. Epub 2019 Sep 11.
We tested whether a mechanical device (such as Hipsecure) to pinpoint the anterior pelvic plane (APP) as a guide can improve acetabular cup placement. To assess accuracy we asked: (1) is the APP an effective guide to position acetabular cup placement within acceptable ° of divergence from the optimal 40° inclination and 15° anteversion; (2) could a mechanical device increase the number of acetabular cup placements within Lewinnek's safe zone (i.e. inclination 30° to 50°; anteversion 5° to 25°)?
16 cadaveric specimens were used to assess the 3D surgical success of using a mechanical device APP to guide acetabular cup placement along the APP. We used the Hipsecure mechanical device to implant acetabular cups at 40° inclination and 15° anteversion. Subequently, all cadaveric specimens with implants were scanned with a CT and 3D models were created of the pelvis and acetabular cups to assess the outcome in terms of Lewinnek's safe zones.
The mean inclination of the 16 implants was 40.6° (95% CI, 37.7-43.4) and the mean anteversion angle was 13.4° (95% CI, 10.7-16.1). All 16 cup placements were within Lewinnek's safe zone for inclination (between 30° and 50°) and all but 2 were within Lewinnek's safe zone for anteversion (between 5° and 25°).
In cadaveric specimens, the use of a mechanical device and the APP as a guide for acetabular cup placement resulted in good positioning with respect to both of Lewinnek's safe zones.
我们测试了一种机械装置(如 Hipsecure)是否可以精确定位前骨盆平面(APP)作为指导,以改善髋臼杯的放置位置。为了评估准确性,我们提出了以下问题:(1)APP 是否可以有效地指导髋臼杯的放置位置,使其与最佳 40°倾斜度和 15°前倾角的偏差在可接受的范围内;(2)机械装置是否可以增加髋臼杯放置在 Lewinnek 安全区内的数量(即倾斜度 30°至 50°;前倾角 5°至 25°)?
使用 16 个尸体标本评估使用机械装置 APP 沿 APP 引导髋臼杯放置的 3D 手术成功率。我们使用 Hipsecure 机械装置以 40°倾斜度和 15°前倾角植入髋臼杯。随后,对所有带有植入物的尸体标本进行 CT 扫描,并创建骨盆和髋臼杯的 3D 模型,以评估 Lewinnek 安全区的结果。
16 个植入物的平均倾斜度为 40.6°(95%置信区间,37.7-43.4),平均前倾角为 13.4°(95%置信区间,10.7-16.1)。所有 16 个杯状植入物均位于 Lewinnek 安全区内的倾斜度(30°至 50°之间),且除 2 个外,所有植入物均位于 Lewinnek 安全区内的前倾角(5°至 25°之间)。
在尸体标本中,使用机械装置和 APP 作为髋臼杯放置的指导,在Lewinnek 安全区的两个方面都实现了良好的定位。