Computed tomography (CT) is the gold standard for the accurate measurement of cup inclination and anteversion in total hip arthroplasty (THA). However, functional cup position strongly depends on the tilt of the pelvis and thus pelvic tilt in the supine position on the CT table harbors the risk of misinterpretation.
To evaluate the influence of pelvic tilt on cup measurements in postoperative CT scans.
This is a secondary outcome analysis of a prospective study. In 123 patients undergoing THA, anatomic cup inclination and anteversion as measured on postoperative 3D-CT scans were compared between: (i) the anterior pelvic plane (APP) resembling neutral pelvic tilt; and (ii) the coronal plane representing the standard planes on CT. Furthermore, the effect of the variation on cup positions within Lewinnek's safe zone in relation to the applied reference plane was assessed.
Mean pelvic tilt in supine position was -3.5° ± 5.6°. This resulted in a mean difference for anteversion between APP und coronal plane of 2.1° ± 3.7° and of 2.1° ± 1.9° for cup inclination, respectively. The change varied up to 20.2° for cup inclination and up 12.2° for cup anteversion. The overall conversion rate for inclination and/or anteversion regarding Lewinnek's safe zone was high at 23.6% (29/123).
Pelvic tilt affects cup orientation on CT scans depending on the applied reference plane. Standard CT planes should be corrected regarding pelvic tilt before measurements to avoid errors in measurement.
计算机断层扫描(CT)是测量全髋关节置换术(THA)中杯倾斜和前倾角的金标准。然而,功能杯位置强烈依赖于骨盆的倾斜,因此 CT 台上仰卧位的骨盆倾斜存在误判的风险。
评估骨盆倾斜对术后 CT 扫描中杯测量的影响。
这是一项前瞻性研究的二次结果分析。在 123 例接受 THA 的患者中,比较了术后 3D-CT 扫描中解剖杯倾斜和前倾角:(i)类似于中立骨盆倾斜的前骨盆平面(APP);和(ii)代表 CT 标准平面的冠状面。此外,还评估了在应用参考平面时,Lewinnek 安全区内杯位置的变化对杯位置的影响。
仰卧位时平均骨盆倾斜为-3.5°±5.6°。这导致 APP 与冠状面之间的前倾角平均差异为 2.1°±3.7°,杯倾斜的平均差异为 2.1°±1.9°。变化幅度最大可达 20.2°,杯倾斜可达 12.2°。根据 Lewinnek 安全区,倾斜和/或前倾角的总体转换率为 23.6%(29/123)。
骨盆倾斜会影响 CT 扫描上的杯定向,具体取决于应用的参考平面。在进行测量之前,应根据骨盆倾斜校正标准 CT 平面,以避免测量误差。