Rajanbabu Balram Babu, Dhandapani Kalaimani
Department of Paediatric Cardiac Surgery, Apollo Hospitals, Hyderabad, India; and.
Department of Clinical Perfusion, Narayana Hrudayalaya Institute of Cardiac Sciences, Bangalore, India.
J Extra Corpor Technol. 2019 Sep;51(3):169-171.
Perfusion-related problems are still reported at incidence, varying from .72 to 2.86%, and add to the morbidity and mortality of cardiac surgery. Awareness of such problems and establishing protocols to avoid and manage them is a key factor in further improving patient outcomes. In this report, we describe a homemade setup which was used as a desperate salvage measure to rewarm a 1-year-old baby on cardiopulmonary bypass for intracardiac repair, who was inadvertently cooled to 24°C because of a malfunctioning heater/cooler machine.
仍有关于灌注相关问题的发生率报告,范围从0.72%至2.86%不等,这些问题增加了心脏手术的发病率和死亡率。意识到这些问题并制定避免和处理这些问题的方案是进一步改善患者预后的关键因素。在本报告中,我们描述了一种自制装置,该装置被用作一种绝望的挽救措施,用于在体外循环下心内修复手术中为一名1岁婴儿复温,该婴儿因加热器/冷却器机器故障意外被冷却至24°C。