Lucchini Alberto, Bambi Stefano, Elli Stefano, Bruno Martina, Dallari Roberto, Puccio Pamela, Villa Silvia, Rona Roberto, Fumagalli Roberto, Bellani Giacomo, Foti Giuseppe
ASST Monza, Ospedale San Gerardo - università degli Studi di Milano-Bicocca.
Acta Biomed. 2019 Nov 11;90(11-S):65-71. doi: 10.23750/abm.v90i11-S.8820.
During Continuous Positive Pressure Ventilation delivered through helmet, the patient inhales high flows of gas without adequate conditioning. However, the need to humidify the inspired gas during Helmet-CPAP, has not been sufficiently explored.
Experimental design study. Six healthy individuals underwent High Flow Helmet CPAP with different gas flows (60 and 80 L/min) and FiO2 (0.35, 0.5, 0.7 and 1) generated by a Venturi system, with and without active humidification. The active humidifier setting was 26 ° at the humidification chamber and 28 ° at the helmet gas inlet. At each setting, measurements about temperature and relative humidity inside helmet were taken. Comfort level at each setting was evaluated using a visual analog scale rated from 0 to 10.
Without heated humidification, the mean value of absolute humidity in the eight combinations investigated was 5.9±2.1 mg H2O/L, with a mean temperature of 25.8±0.9°. With heated humidification mean absolute humidity was 15.0±3.5 mg H2O/L with mean temperature of 29.0±0.1°. The median comfort scale value was 6 (IQR: 5.25-6.75) during the phase without humidification vs 8 (7.25-8.0 - P<0.01) when active humidification was applied.
In healthy subjects undergoing High Flow Helmet CPAP, heated humidifiers with heated wires tubes are necessary to avoid the under-humidification inside the helmet. To obtain patient's comfort and airways mucosal humidification during continuous Helmet CPAP, the most desirable conditions are reached by heated humidifiers with a humidifying chamber temperature settled between 26-28°C.
在通过头盔进行持续气道正压通气期间,患者吸入的高流量气体未得到充分调节。然而,头盔式持续气道正压通气期间对吸入气体进行加湿的必要性尚未得到充分研究。
实验设计研究。六名健康个体接受了由文丘里系统产生的不同气体流量(60和80升/分钟)和氧浓度(0.35、0.5、0.7和1)的高流量头盔式持续气道正压通气,分为有和没有主动加湿两种情况。主动加湿器在加湿腔的设置温度为26°C,在头盔气体入口处为28°C。在每种设置下,测量头盔内的温度和相对湿度。使用0至10的视觉模拟量表评估每种设置下的舒适度。
在没有加热加湿的情况下,所研究的八种组合中绝对湿度的平均值为5.9±2.1毫克水/升,平均温度为25.8±0.9°C。在有加热加湿的情况下,平均绝对湿度为15.0±3.5毫克水/升,平均温度为29.0±0.1°C。在未加湿阶段,舒适度量表的中位数为6(四分位距:5.25 - 6.75),而在应用主动加湿时为8(7.25 - 8.0 - P<0.01)。
在接受高流量头盔式持续气道正压通气的健康受试者中,带有加热丝管的加热加湿器对于避免头盔内加湿不足是必要的。为了在持续头盔式持续气道正压通气期间获得患者的舒适度和气道黏膜加湿,加湿腔温度设定在26 - 28°C之间的加热加湿器能达到最理想的条件。