MAGMA. 2020 Aug;33(4):549-557. doi: 10.1007/s10334-019-00804-9. Epub 2019 Nov 28.
Synthetic magnetic resonance imaging (SyMRI) allows to obtain different weighted-images using the multiple-dynamic multiple-echo sequence lasting 6 min. The aim is to compare quantitatively and qualitatively synthetic- and conventional MRI in patients with multiple sclerosis (MS) and controls assessing the contrast (C), the signal to noise ratio (SNR), and the contrast to noise ratio (CNR). We evaluated the lesion count and lesion-to-white matter contrast ([Formula: see text] in the MS patients.
51 patients underwent synthetic- and conventional MRI. Qualitative analysis was evaluated by assigning scores to all synthetic- and conventional MRI sequences by two neuroradiologists. Lesions were counted in MS patients both in the conventional- and synthetic T2-FLAIR. Regions of interest were placed in the cerebrospinal fluid, in the white- and grey matter. For the sequences were evaluated: C, CNR, and SNR.
Synthetic T2-FLAIR images are qualitatively inferior. C and CNR were significantly higher in synthetic T1W and T2W images compared to conventional images, but not for T2-FLAIR. The SNR value was always lower in synthetic images than in conventional ones.
SyMRI can be used in clinical practice because it has a similar diagnostic accuracy which reduces the scanning time compared to the conventional one. However, synthetic T2-FLAIR images need to be improved.
合成磁共振成像(SyMRI)可以使用持续 6 分钟的多动态多回波序列获得不同加权图像。目的是通过定量和定性比较多发性硬化症(MS)患者和对照组的合成和常规 MRI,评估对比(C)、信噪比(SNR)和对比噪声比(CNR)。我们评估了 MS 患者的病变计数和病变与白质的对比[公式:见文本]。
51 例患者接受了合成和常规 MRI 检查。两名神经放射科医生对所有合成和常规 MRI 序列进行评分,对定性分析进行评估。在常规 T2-FLAIR 和合成 T2-FLAIR 中对 MS 患者的病变进行计数。在脑脊液、白质和灰质中放置感兴趣区。评估了序列:C、CNR 和 SNR。
合成 T2-FLAIR 图像质量较差。与常规图像相比,合成 T1W 和 T2W 图像的 C 和 CNR 明显更高,但 T2-FLAIR 图像则不然。与常规图像相比,合成图像的 SNR 值始终较低。
SyMRI 可在临床实践中使用,因为它具有相似的诊断准确性,与常规 MRI 相比,可缩短扫描时间。然而,合成 T2-FLAIR 图像需要改进。